1.
Int J Artif Organs ; 21(3): 147-50, 1998 Mar.
Article in English | MEDLINE | ID: mdl-9622113

ABSTRACT

This study gives the results in terms of precision and repeatability of a new on-line urea monitoring system (Ureascan P2 Hospal) capable of measuring the urea concentrations in the spent dialysate. The Ureascan P2 Hospal (UP2H), fitted on single-pass dialysis machines (Integra-Hospal), functions by the presence of a disposable mini-reactor containing urease. The passage through the reactor of a minimum quantity of spent dialysate from the filter diluted with a pH 7 buffer solution (1 ml/min) increases its ionic strength, which is detected by a differential measurement of conductivity in proportion to the urea concentration in the dialysis liquid. We studied 13 dialysis sessions, with bicarbonate buffer, in 8 anuric patients. From 4 to 7 dialysate samples were taken during each treatment to determine the urea and 65 samples were analysed overall. Urea values from the UP2H were compared with those measured on the Dimension Du Pont analyser. Simple linear regression analysis showed an excellent correlation between the 2 measuring methods (r=0.987; p<0.001). The Bland-Altman test gave an average difference between the urea values measured with the UP2H and in the laboratory of 1.3+/-1.2 mg/dl. The agreement limits between 2 SD were -1.2 mg/dl and +3.8 mg/dl respectively. In conclusion, the UP2H we have developed has proved to be a reliable and very useful instrument for adapting, through the urea kinetic mathematical models, the dialysis dose for individual patients.


Subject(s)
Biosensing Techniques , Dialysis Solutions/analysis , Renal Dialysis , Urea/analysis , Anuria/therapy , Bicarbonates/chemistry , Humans , Models, Theoretical , Reproducibility of Results , Urease/chemistry
SELECTION OF CITATIONS
SEARCH DETAIL
...