Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 3 de 3
Filter
Add more filters










Database
Language
Publication year range
1.
Europace ; 1(3): 174-8, 1999 Jul.
Article in English | MEDLINE | ID: mdl-11225793

ABSTRACT

AIM: A prospective randomized trial was set up to evaluate contractile parameters and quality of life in patients with congestive heart failure. METHODS AND RESULTS: We describe the results from 38 patients in sinus rhythm and with chronic heart failure due to congestive cardiomyopathy, prospectively randomized to optimal medical therapy (Group 1, 19 patients) or optimal medical therapy plus dual chamber pacemaker programmed to optimal AV delay (Group 2, 19 patients). At a 6 month follow-up, 7/19 patients in Group 1 had died compared with 5/19 patients in Group 2. During follow-up, there were few significant changes in evaluated parameters except for mitral regurgitation time, which was prolonged in Group 1 and shortened in Group 2. The systolic left ventricular diameter shortened significantly only in Group 2. An energy and activity questionnaire showed that the effect of DDD pacing in the latter patient population was beneficial. CONCLUSIONS: From these results we may conclude that at the 6 month follow-up DDD pacing with echo-optimized AV interval programming can improve quality of life without affecting survival.


Subject(s)
Heart Failure/therapy , Pacemaker, Artificial , Aged , Aged, 80 and over , Equipment Design , Female , Heart Failure/physiopathology , Humans , Male , Middle Aged , Time Factors
2.
Pacing Clin Electrophysiol ; 19(11 Pt 2): 1719-23, 1996 Nov.
Article in English | MEDLINE | ID: mdl-8945030

ABSTRACT

To evaluate the safety and efficacy of a new algorithm for automatic mode switching (AMS) from DDD-DDDR to DDIR, 26 patients, 16 females and 10 males, mean age 73 +/- 6 years of age, affected by sinus node disease, chronotropic incompetence, and recurrent paroxysmal atrial fibrillation (PAF) received the Medtronic Thera DR pacemaker. The device continuously calculates, in ms, the running average of the intrinsic atrial rate (MAR) and compares the current atrial interval (CAI) with the stored MAR. When the CAI is greater than the MAR it increases by 8 ms, and when the CAI is less than the MAR, it decreases by 23 ms. When MAR < or = 330 ms (182 beats/min), tachycardia is detected and AMS is activated. All patients had clinical evaluation, 12-lead ECG, Holter monitoring, and exercise testing after implantation and every 3 months for 1 year. The results were compared with the data stored in the pacemaker memory: AMS episodes number; the histogram of the last 14 episodes; and atrial electrogram recording. Twenty-two Holter recordings in 13 patients showed PAF and in all of them AMS occurred simultaneously. AMS lasted between 10 seconds and 20 hours, and MAR ranged from 195-400 beats/min. No episode of PAF and no AMS were recorded in 39 Holter recordings in 22 patients. Appropriate AMS was confirmed in five patients by stored atrial electrogram and in nine by 12-lead ECG and pacemaker event markers. Mean atrial sensing was 2.13 +/- 1.04 mV during PAF and 3.18 +/- 1.46 mV during sinus rhythm. No PAF episode and no AMS were recorded during exercise testing. In conclusion, this new algorithm was very reliable, sensitive, and specific.


Subject(s)
Algorithms , Atrial Fibrillation/therapy , Cardiac Pacing, Artificial/methods , Heart Rate , Sick Sinus Syndrome/therapy , Aged , Atrial Fibrillation/physiopathology , Atrial Function , Electrocardiography/methods , Electrocardiography, Ambulatory , Equipment Design , Equipment Safety , Exercise Test , Female , Follow-Up Studies , Humans , Male , Pacemaker, Artificial , Recurrence , Reproducibility of Results , Sensitivity and Specificity , Sick Sinus Syndrome/physiopathology , Software , Tachycardia/diagnosis , Tachycardia/therapy , Time Factors
3.
J Steroid Biochem Mol Biol ; 51(5-6): 319-22, 1994 Dec.
Article in English | MEDLINE | ID: mdl-7826895

ABSTRACT

The existence of a specific binding site for sex steroid binding protein (SBP or SHBG) was detected on plasma membranes prepared from the testis of a patient affected by a variant form of testicular feminization. A binding technique using [125I]SBP as a tracer allowed us to identify a single set of binding sites, characterized by a Kd of 1.917 x 10(-11) M. The maximum number of binding sites was 5.2 fmol/mg membrane protein. Membranes were also prepared from a sample of genital skin from the same patient, but no binding for [125I]SBP was detectable. The evidence of the SBP membrane receptor in the testis of a patient affected by Morris syndrome extends our knowledge about the tissue distribution of the SBP receptor and suggests the possible implication of SBP and its recognition system in a disorder related to peripheral androgen insensitivity.


Subject(s)
Androgen-Insensitivity Syndrome/metabolism , Receptors, Cell Surface/metabolism , Sex Hormone-Binding Globulin/metabolism , Testis/metabolism , Cell Membrane/metabolism , Genitalia, Male/metabolism , Humans , Male , Middle Aged , Protein Binding , Radioligand Assay , Skin/metabolism
SELECTION OF CITATIONS
SEARCH DETAIL
...