1.
Pacing Clin Electrophysiol ; 37(7): 889-99, 2014 Jul.
Article in English | MEDLINE | ID: mdl-24527748

ABSTRACT

INTRODUCTION: Adjudication of thousands of implantable cardioverter defibrillator (ICD)-treated arrhythmia episodes is labor intensive and, as a result, is most often left undone. The objective of this study was to evaluate an automatic classification algorithm for adjudication of ICD-treated arrhythmia episodes. METHODS: The algorithm uses a machine learning algorithm and was developed using 776 arrhythmia episodes. The algorithm was validated on 131 dual-chamber ICD shock episodes from 127 patients adjudicated by seven electrophysiologists (EPs). Episodes were classified by panel consensus as ventricular tachycardia/ventricular fibrillation (VT/VF) or non-VT/VF, with the resulting classifications used as the reference. Subsequently, each episode electrogram (EGM) data was randomly assigned to three EPs without the atrial lead information, and to three EPs with the atrial lead information. Those episodes were also classified by the automatic algorithm with and without atrial information. Agreement with the reference was compared between the three EPs consensus group and the algorithm. RESULTS: The overall agreement with the reference was similar between three-EP consensus and the algorithm for both with atrial EGM (94% vs 95%, P = 0.87) and without atrial EGM (90% vs 91%, P = 0.91). The odds of accurate adjudication, after adjusting for covariates, did not significantly differ between the algorithm and EP consensus (odds ratio 1.02, 95% confidence interval: 0.97-1.06). CONCLUSIONS: This algorithm performs at a level comparable to an EP panel in the adjudication of arrhythmia episodes treated by both dual- and single-chamber ICDs. This type of algorithm has the potential for automated analysis of clinical ICD episodes, and adjudication of EGMs for research studies and quality analyses.


Subject(s)
Algorithms , Arrhythmias, Cardiac/classification , Arrhythmias, Cardiac/physiopathology , Defibrillators, Implantable , Electrophysiologic Techniques, Cardiac , Humans
2.
Heart Rhythm ; 10(5): 702-8, 2013 May.
Article in English | MEDLINE | ID: mdl-23337541

ABSTRACT

BACKGROUND: The efficacy of shock in converting different ventricular tachyarrhythmias has not been well characterized in a large natural-practice setting. OBJECTIVE: To determine shock success rate by energy and ventricular rhythm in a large cohort of patients with implantable cardioverter-defibrillators. METHODS: Two thousand patients with 5279 shock episodes were randomly sampled for analysis from the LATITUDE remote monitoring system. Within an episode, the rhythm preceding therapy (shock or antitachycardia pacing [ATP]) was adjudicated. Patients who died after unsuccessful implantable cardioverter-defibrillator shocks did not transmit final remote monitoring data and were not included in the study. RESULTS: Of 3677 shock episodes for ventricular tachyarrhythmia, 2679 were treated with shock initially and were classified as monomorphic ventricular tachycardia ( n = 1544), polymorphic/monomorphic ventricular tachycardia (n = 371), or ventricular fibrillation (n = 764). The success rate after the first, second, and final shock averaged 90.3%, 96.4%, and 99.8%, respectively. After unsuccessful initial ATP (n = 998), the first, second, and final shock was successful in 84.8%, 92.9%, and 100% of the episodes. The success rate after the first or second shock was significantly lower after failed ATP compared to shock as first therapy (both P<.001). Among episodes treated initially with shock, the success rate for monomorphic ventricular tachycardia (89.2%) when treated with energy level ≤ 20 J was significantly higher than that for ventricular fibrillation (80.8%) (P = .04). The level of shock energy was a significant predictor of the success of the first shock (odds ratio 1.16; 95% confidence interval 1.03-1.30; P = .013). CONCLUSIONS: The success rate of first shock as first therapy is approximately 90%, but was lower after failed ATP. Programming a higher level of energy after ATP is suggested.


Subject(s)
Defibrillators, Implantable , Electric Countershock , Heart Ventricles/physiopathology , Tachycardia, Ventricular/physiopathology , Ventricular Fibrillation/physiopathology , Aged , Aged, 80 and over , Electrocardiography , Female , Humans , Male , Middle Aged , Survival Rate , Tachycardia, Ventricular/therapy , Treatment Outcome , United States , Ventricular Fibrillation/therapy
3.
Pacing Clin Electrophysiol ; 35(7): 863-9, 2012 Jul.
Article in English | MEDLINE | ID: mdl-22519674

ABSTRACT

BACKGROUND: Approximately 12-21% of implantable cardioverter defibrillator (ICD) patients receive inappropriate shocks. We sought to determine the incidence and causes of noise/artifact and oversensing (NAO) resulting in ICD shocks. METHODS: A random sample of 2,000 patients who received ICD and cardiac resynchronization therapy defibrillator shocks and were followed by a remote monitoring system was included. Seven electrophysiologists analyzed stored electrograms from the 5,279 shock episodes. Episodes were adjudicated as appropriate or inappropriate shocks. RESULTS: Of the 5,248 shock episodes with complete adjudication, 1,570 (30%) were judged to be inappropriate shocks. Of these 1,570, 134 (8.5%) were a result of NAO. The 134 NAO episodes were determined to be due to external noise in 76 (57%), lead connector-related in 37 (28%), muscle noise in 11 (8%), oversensing of atrium in seven (5%), T-wave oversensing in two (2%), and other noise in one (1%). The ICD shock itself resulted in a marked decrease in the level of noise in 60 of 134 (45%) NAO episodes, and the magnitude of this effect varied with the type of NAO (58% for external noise, 35% for muscle, 27% for lead/connector, and 0% for oversensing; P = 0.03). There was no significant difference in NAO likelihood based on type of lead (integrated bipolar 89/1,802 vs dedicated bipolar 9/140, P = 0.67). CONCLUSIONS: External noise and lead/connector noise were the primary causes, while T-wave oversensing was the least common cause of NAO resulting in ICD shock. Noise/artifact decreased immediately after a shock in nearly half of episodes. The specific ICD lead type did not impact the likelihood of NAO.


Subject(s)
Artifacts , Defibrillators, Implantable/statistics & numerical data , Electric Injuries/epidemiology , Equipment Failure/statistics & numerical data , Signal-To-Noise Ratio , Female , Humans , Incidence , Male , Risk Factors , United States/epidemiology
4.
Surg Today ; 42(8): 793-6, 2012 Aug.
Article in English | MEDLINE | ID: mdl-22361987

ABSTRACT

The intra-aortic balloon pump (IABP) can be used transiently to improve cardiac function mechanically, in patients with severe cardiomyopathy and heart failure refractory to medical therapy. In the field of surgery, the IABP is most commonly used for patients with myocardial infarction, congestive heart failure, or other chronic cardiac conditions, who are undergoing cardiac surgery. Conversely, it is rarely used in hepatobiliary surgery, with only two reports found in the literature, excluding cases of emergency cholecystectomy. We describe how we used an IABP successfully during surgery to repair a transected bile duct in a patient with peripartum cardiomyopathy.


Subject(s)
Bile Duct Diseases/surgery , Bile Ducts, Intrahepatic/surgery , Cardiomegaly/complications , Cardiomyopathies/complications , Intra-Aortic Balloon Pumping , Puerperal Disorders/surgery , Adult , Bile Duct Diseases/complications , Bile Duct Diseases/diagnosis , Bile Ducts, Intrahepatic/pathology , Cardiomegaly/diagnosis , Cardiomyopathies/diagnosis , Female , Humans , Puerperal Disorders/diagnosis
5.
World J Pediatr Congenit Heart Surg ; 3(1): 123-9, 2012 Jan 01.
Article in English | MEDLINE | ID: mdl-23804696

ABSTRACT

Submitted May 6, 2011; Accepted August 3, 2011. The survival into adulthood of patients with unoperated complex congenital heart disease with anomalies often considered life threatening in infancy and childhood requires a complex interplay of "balanced" defects allowing for cardiovascular physiology compatible with long-term survival. We report on a series of three cases from our advanced imaging database of middle-aged adults presenting with multiple similar defects providing a hemodynamically balanced circulation. The constellation of defects seen in each of these patients included congenitally corrected transposition of the great arteries, a large nonrestrictive ventricular septal defect, valvular pulmonary stenosis, and in two cases anomalous coronary arteries. Cardiovascular computed tomographic angiography (CCTA) and cardiovascular magnetic resonance imaging (CMR) were important to the characterization of the multiple defects and their three-dimensional relationships in these cases. Treatment decisions in patients with this constellation of findings are challenging, given the limited data due to the rarity of survival of patients with these defects into middle adulthood and the paucity of data related to decisions and approaches to medical management, surgical correction, or transplantation.

6.
Pacing Clin Electrophysiol ; 34(7): 821-6, 2011 Jul.
Article in English | MEDLINE | ID: mdl-21535040

ABSTRACT

BACKGROUND: Electrophysiology procedures vary in invasiveness, duration, and anesthesia utilized. While complications are low and efficacy high, cases are elective and patient experiences related to anxiety, pain, and perceived outcomes are not well studied. We sought to determine if a 30-minute audio compact disc (CD) that teaches relaxation techniques and wellness perception prior to an elective procedure impacts validated measures of anxiety, pain, and procedural outcomes. METHODS: Sixty-one patients were randomly assigned to a control group (CG) (N(CG) = 31) or interventional group (IG) (N(IG) = 30). Both groups answered a baseline Hospital Anxiety and Depression Scale (HADS-A) survey consisting only of anxiety assessment questions. The IG listened to the CD the night prior to their procedure. Heart rate and blood pressure were monitored on admission and prior to the procedure. Postprocedure, both groups completed two HADS-A surveys as well as two Patient Experience Surveys (PES). There was no statistical difference in the demographics and the rate of procedural complications between the groups. The statistical significance of our data was determined using a Student's t-test and χ(2) test. RESULTS: At baseline, both groups had equal amounts of anxiety prior to their procedures (P = 0.2). The patients in the IG had lower systolic blood pressures during admission and prior the administration of analgesics in comparison to the CG. Postprocedure, results from administering the HADS-A demonstrated that the IG had 33% lower anxiety (P = 0.02) than CG patients. CONCLUSION: The implementation of basic relaxation teaching techniques prior to planned electrophysiology procedures lowers systolic blood pressure and postprocedural anxiety.


Subject(s)
Anxiety/etiology , Anxiety/prevention & control , Electrophysiologic Techniques, Cardiac/adverse effects , Pain/etiology , Pain/prevention & control , Relaxation Therapy , Adolescent , Adult , Aged , Aged, 80 and over , Female , Humans , Male , Middle Aged , Surveys and Questionnaires , Treatment Outcome , Young Adult
7.
Pacing Clin Electrophysiol ; 34(8): 1003-12, 2011 Aug.
Article in English | MEDLINE | ID: mdl-21453341

ABSTRACT

BACKGROUND: The increasing use of remote monitoring with the associated large retrievable databases provides a unique opportunity to analyze observations on implantable cardioverter-defibrillator (ICD) therapies. Adjudication of a large number of stored ICD electrograms (EGMs) presents a unique challenge. The ALTITUDE study group was designed to use the LATITUDE remote monitoring system to evaluate ICD patient outcomes across the United States. METHODS AND RESULTS: Of 81,081 patients on remote monitoring, a random sample of 2,000 patients having 5,279 shock episodes was selected. The ALTITUDE EGM review committee was comprised of seven electrophysiologists from four institutions. An online EGM adjudication system was designed. Episodes were classified as appropriate (70% of shock episodes) or inappropriate ICD therapies (30%). Light's Kappa was used to assess agreement. Interobserver and intraobserver Kappa scores for dual-chamber ICDs were 0.84 (0.71-0.91) and 0.89 (0.82-0.95), consistent with substantial agreement. Interobserver and intraobserver Kappa scores for single-chamber ICDs were 0.61 (0.54-0.67) and 0.69 (0.59-0.79). The rhythm categories of "nonsustained arrhythmia" and "polymorphic and monomorphic ventricular tachycardia" resulted in the greatest degree of discordant adjudication between reviewers. CONCLUSIONS: This method of adjudication of a large volume of stored EGM data prior to device therapies will allow new observations in regards to device performance and has the potential to improve device programming and design. There was substantial interreviewer agreement for rhythm classification. Agreement was greater for dual-chamber compared to single-chamber devices, indicating the atrial lead adds diagnostic value in rhythm interpretation.


Subject(s)
Defibrillators, Implantable , Electrophysiologic Techniques, Cardiac , Monitoring, Physiologic/methods , Arrhythmias, Cardiac/diagnosis , Arrhythmias, Cardiac/therapy , Humans , Registries , Tachycardia, Ventricular/diagnosis , Tachycardia, Ventricular/therapy , Treatment Outcome
8.
Curr Cardiol Rep ; 13(3): 203-9, 2011 Jun.
Article in English | MEDLINE | ID: mdl-21445560

ABSTRACT

Heart failure is a major public health concern that is frequently complicated by ventricular arrhythmias. Sustained ventricular tachycardia is associated with an increased risk for progressive heart failure and sudden death. We summarize the current management strategies for ventricular tachycardia in heart failure patients, including implantable cardioverter-defibrillator therapy, pharmacologic therapy, catheter ablation techniques, ventricular assist device therapy, and heart transplantation.


Subject(s)
Heart Failure/complications , Tachycardia, Ventricular/etiology , Tachycardia, Ventricular/therapy , Anti-Arrhythmia Agents/pharmacology , Catheter Ablation , Defibrillators, Implantable , Heart Transplantation , Humans , Randomized Controlled Trials as Topic
9.
J Cardiovasc Electrophysiol ; 22(3): 359-63, 2011 Mar.
Article in English | MEDLINE | ID: mdl-20958827

ABSTRACT

Sustained ventricular tachycardia (VT) in patients with advanced cardiomyopathy is a potentially life-threatening arrhythmia. Newer treatment strategies have evolved that combine the use of catheter ablation to target the substrate for VT and ventricular assist devices (VADs) to hemodynamically support the failing ventricle. This editorial is targeted to the practicing clinician caring for these difficult patients. The current article reviews the use of percutaneous VADs to support catheter ablation of VT, the use of durable VADs to support the failing heart in patients with recurrent VT, ventricular arrhythmias in patients with durable VADs, and the use of catheter ablation to treat VT in patients with durable VADs.


Subject(s)
Cardiomyopathies/complications , Catheter Ablation , Heart Failure/therapy , Heart-Assist Devices , Tachycardia, Ventricular/therapy , Cardiomyopathies/physiopathology , Heart Failure/etiology , Heart Failure/physiopathology , Hemodynamics , Humans , Radiography, Interventional , Recurrence , Tachycardia, Ventricular/etiology , Tachycardia, Ventricular/physiopathology , Treatment Outcome , Ventricular Function
10.
J Am Soc Echocardiogr ; 23(10): 1113.e1-4, 2010 Oct.
Article in English | MEDLINE | ID: mdl-20434878

ABSTRACT

The authors describe the case of a 79-year-old man with prior mitral valve repair and a maze procedure who developed recurrent atrial fibrillation, in whom transesophageal echocardiography revealed an accessory lobe of the left atrial appendage in sinus rhythm when the remaining body of the left atrial appendage was in atrial fibrillation or flutter. Electrophysiology confirmed dissociated rhythm within the left atrium. This case emphasizes the need for careful Doppler interrogation of the left atrial appendage and its lobes to look for dissociated atrial rhythm.


Subject(s)
Atrial Appendage/diagnostic imaging , Atrial Fibrillation/diagnostic imaging , Echocardiography, Transesophageal/methods , Heart Atria/diagnostic imaging , Aged , Humans , Male
11.
J Cardiovasc Electrophysiol ; 21(1): 99-102, 2010 Jan.
Article in English | MEDLINE | ID: mdl-19549034

ABSTRACT

We describe the case of a young patient with severe hypertrophic cardiomyopathy and marginal defibrillation thresholds (DFTs) at implant of a standard transvenous implantable cardioverter-defibrillator (ICD) system. The patient subsequently experienced multiple failed ICD shocks during a prolonged episode of spontaneous ventricular tachycardia/fibrillation. Placement of a second single-coil shocking lead in the azygous vein resulted in acceptable DFTs, but the new lead migrated superiorly within hours of the procedure. To stabilize the lead position, a vascular plug was placed in the distal azygous vein, and the shocking lead screw was actively fixated to the meshwork of the device. Subsequent testing confirmed both adequate defibrillation and stable lead position.


Subject(s)
Blood Vessel Prosthesis , Cardiomyopathies/prevention & control , Cardiomyopathy, Hypertrophic/prevention & control , Defibrillators, Implantable , Electrodes, Implanted , Prosthesis Implantation/instrumentation , Prosthesis Implantation/methods , Adolescent , Cardiomyopathy, Hypertrophic/diagnosis , Female , Humans , Treatment Outcome
12.
Pacing Clin Electrophysiol ; 32(3): 340-5, 2009 Mar.
Article in English | MEDLINE | ID: mdl-19272064

ABSTRACT

BACKGROUND: Device extraction is a critical component in the treatment of patients with device-related infections. Due to complex anatomic considerations, device extraction in adults with congenital heart disease presents unique challenges to the electrophysiologist. METHODS: Here, we present a series of device-extraction cases performed in patients with transposition of the great arteries status post either Mustard or Senning surgical procedures that subsequently had permanent pacemakers placed and ultimately developed device-related infections. RESULTS: All of these patients eventually underwent successful laser extractions of their infected devices resulting in complete removal of all hardware and resolution of their infections. CONCLUSIONS: These cases illustrate that endovascular device extraction has been safely and effectively performed in adult patients with congenital heart disease, though further studies are needed to determine the procedural risks and success rates of this procedure in this patient population.


Subject(s)
Device Removal/instrumentation , Device Removal/methods , Electrodes, Implanted/adverse effects , Heart Defects, Congenital/surgery , Prosthesis-Related Infections/prevention & control , Adult , Female , Humans , Male , Treatment Outcome
13.
J Am Coll Cardiol ; 51(23): 2241-9, 2008 Jun 10.
Article in English | MEDLINE | ID: mdl-18534271

ABSTRACT

OBJECTIVES: The purpose of this study was to define the incidence, mechanisms, and management, including catheter ablation, of supraventricular tachycardia (SVT) in a large series of patients after orthotopic heart transplantation (OHT). BACKGROUND: Supraventricular arrhythmias are frequently encountered after OHT, but their characteristics in this population have not been well established. METHODS: We analyzed the incidence, clinical course, and management of SVTs in a cohort of 729 adult patients who underwent OHT. Furthermore, the mechanisms of arrhythmias among the patients referred for electrophysiological study (EPS) and ablation were also characterized. RESULTS: The most common arrhythmia was atrial flutter, which occurred in 9% of this cohort. Persistent or paroxysmal atrial fibrillation occurred in 7%, the majority (57%) in the perioperative period. Persistent or paroxysmal atrial fibrillation was observed in OHT patients, beyond the post-operative period, only in the presence of rejection or transplant vasculopathy. Other persistent or paroxysmal SVTs were seen in 47 stable OHT patients (7%). Of these, 24 patients (4%) underwent EPS. Accessory and dual atrioventricular nodal pathways in the donor heart caused SVT in 3 patients. Macro-reentrant atrial tachycardia was seen in 7 patients, and isthmus-dependent atrial flutter occurred in 14 patients. CONCLUSIONS: The majority of SVTs in stable OHT patients can be attributed to macro-reentrant tachycardias (flutter and scar reentry). Catheter ablation is effective in management of these SVTs. Atrial fibrillation was never encountered in stable patients in our series, and its occurrence should prompt an evaluation for acute rejection and/or vasculopathy.


Subject(s)
Catheter Ablation , Heart Transplantation/adverse effects , Tachycardia, Supraventricular/etiology , Atrial Fibrillation/etiology , Female , Heart Rate , Heart Transplantation/methods , Humans , Incidence , Male , Middle Aged , Retrospective Studies , Risk Factors , Tachycardia, Supraventricular/mortality , Tachycardia, Supraventricular/physiopathology , Tachycardia, Supraventricular/therapy , Time Factors
14.
J Interv Card Electrophysiol ; 21(3): 223-6, 2008 Apr.
Article in English | MEDLINE | ID: mdl-18297382

ABSTRACT

We present a 46-year-old patient who suffered from cardiac arrest and subsequently underwent placement of an implantable cardioverter defibrillator (ICD). The patient underwent a cardiac catheterization which revealed no significant coronary artery disease. About 1 year later he experienced appropriated and frequent ICD discharges due to monomorphic ventricular tachycardia (VT) with left bundle branch block morphology. His prodromal symptoms were mild dizziness and lightheadedness with no chest pain. Amiodarone, mexiletine, sotalol and dofetilide as well as ablation of two inducible ventricular tachycardias in the electrophysiology studies were unsuccessful in controlling the arrhythmias and ICD discharges. During the last episode, he experienced a mild burning sensation in his chest and was given nitroglycerin 0.4 mg sublingually, which relived his symptoms and aborted the VT. This led to a second cardiac catheterization to investigate whether the VT was being induced by myocardial ischemia. This second coronary angiogram spontaneously revealed significant coronary vasospasm and simultaneously, the patient's cardiac rhythm showed short runs of VT with left bundle branch block morphology. Intracoronary nitroglycerine relieved the coronary vasospasm and terminated the arrhythmia. The patient was treated with isosorbide mononitrate and diltiazem. He remained symptom free with no ICD discharges and no VT in ICD interrogations for more than 2 years. Coronary vasospasm may be silent and with no chest pain which creates a difficult clinical situation particularly if it is associated with ventricular tachycardia and sudden cardiac death. The mechanisms of VT in the setting of coronary vasospasm are not known and increased automaticity, focal discharges, functional unidirectional block with reentry, or a combination of these mechanisms may contribute to inducing the VT during the transient ischemia or rarely in the reperfusion phase. It is important to perform provocative tests to diagnose silent coronary vasospasm in unexplained sudden cardiac arrests.


Subject(s)
Coronary Vasospasm/complications , Tachycardia, Ventricular/etiology , Cardiac Catheterization , Coronary Angiography , Coronary Vasospasm/diagnosis , Coronary Vasospasm/drug therapy , Coronary Vasospasm/physiopathology , Defibrillators, Implantable , Electrocardiography , Humans , Male , Middle Aged , Nitroglycerin/administration & dosage , Tachycardia, Ventricular/physiopathology , Tachycardia, Ventricular/therapy , Vasodilator Agents/administration & dosage
15.
Cardiol Clin ; 25(4): 595-603; vii, 2007 Nov.
Article in English | MEDLINE | ID: mdl-18063163

ABSTRACT

Since the 1970s when the implantable cardioverter defibrillator (ICD) was developed, multiple clinical trials have documented survival benefits in certain high-risk subsets of heart failure patients. Over the past decade, cardiac resynchronization therapy (CRT) emerged as an important therapy in carefully selected patients with ongoing symptoms despite optimized pharmacological therapy. ICDs should be considered first-line therapy for survivors of life-threatening ventricular arrhythmic events. Subsets of patients with both ischemic and nonischemic dilated cardiomyopathy appear to have a survival benefit from primary ICD therapy. CRT has resulted in substantial symptomatic improvement and survival benefits in a subgroup of chronic heart failure patients. CRT should be considered in heart failure patients undergoing ICD implantation who have evidence of ventricular dyssynchrony.


Subject(s)
Cardiac Pacing, Artificial , Defibrillators, Implantable , Heart Failure/therapy , Death, Sudden, Cardiac/prevention & control , Heart Arrest/mortality , Heart Failure/mortality , Heart Failure/physiopathology , Humans , Patient Selection , Practice Guidelines as Topic , Randomized Controlled Trials as Topic , Stroke Volume , Survivors , Tachycardia, Ventricular/therapy , Ventricular Dysfunction, Left/physiopathology
16.
Pacing Clin Electrophysiol ; 30(11): 1363-8, 2007 Nov.
Article in English | MEDLINE | ID: mdl-17976100

ABSTRACT

BACKGROUND: Atrial fibrillation (AF) is commonly associated with heart failure. The benefit of cardiac resynchronization therapy (CRT) on atrial remodeling has been demonstrated. However, biventricular pacing did not reduce the global incidence of AF. We evaluated the relationship between CRT response and AF duration. METHODS: We retrospectively analyzed data from 96 patients (59 +/- 15 years; 78% male) who underwent CRT. All patients had class III-IV New York Heart Association (NYHA) symptoms despite maximal medical therapy, left ventricular ejection fraction (LVEF) < or = 35%, QRS >130 ms, and sinus rhythm before implantation. CRT response in patients who survived at six months of follow-up was defined as: (1) no hospitalization for heart failure and (2) improvement of one or more grades in the NYHA classification. RESULTS: CRT responders (n = 54) and non-responders (n = 42) had similar baseline characteristics, including the incidence of persistent AF within six months before implantation. Six months after implantation, when compared to baseline, CRT responders exhibited a significant decrease in left atrial size (47.5 +/- 7.1 mm vs 44.6 +/- 7.7 mm, P < 0.01) and in the incidence of persistent AF (17% vs 2%, P = 0.02). At six months, CRT responders demonstrated shorter mean AF duration (7.5 +/- 43.3 hours vs 48.8 +/- 129.0 hours, P = 0.03) and lower incidence of persistent AF (2% vs 19%, P = 0.004) compared to nonresponders. CONCLUSION: CRT response is associated with a reversal of atrial remodeling and a shorter AF duration.


Subject(s)
Atrial Fibrillation/diagnosis , Atrial Fibrillation/prevention & control , Cardiac Pacing, Artificial/methods , Female , Humans , Male , Middle Aged , Retrospective Studies , Treatment Outcome
18.
Heart Rhythm ; 4(10): 1274-83, 2007 Oct.
Article in English | MEDLINE | ID: mdl-17905332

ABSTRACT

BACKGROUND: The impact of cardiac resynchronization therapy (CRT) on dispersion of repolarization is controversial. The benefit of CRT on sudden cardiac death has been demonstrated only after 3 years follow-up. OBJECTIVE: The purpose of this study was to explore the immediate effect of CRT on dispersion of repolarization and to define the value of dispersion of repolarization parameters as predictors of appropriate implantable cardioverter-defibrillator (ICD) therapy. METHODS: Data from 100 patients who underwent CRT-ICD placement were analyzed retrospectively. Patients had symptoms of New York Heart Association functional class III or IV heart failure, left ventricular ejection fraction < or =35%, and QRS duration >130 ms or QRS < or =130 ms with left intraventricular dyssynchrony. ECG indices of dispersion of repolarization before and immediately after CRT implantation (QT dispersion, Tpeak-Tend [Tp-e], and Tp-e dispersion) were measured. RESULTS: In patients who were upgraded to a biventricular system, Tp-e did not increase significantly after CRT. However, Tp-e increased significantly after CRT in patients with left bundle branch block or narrow QRS at baseline. After 12-month follow-up, 22 patients had received appropriate ICD therapy. ICD therapy and no ICD therapy groups had similar baseline characteristics, such as secondary prevention and ischemic cardiomyopathy. Postimplantation Tp-e was the only independent predictor of future ICD therapy (P = .02). CONCLUSION: Immediately after CRT, Tp-e did not increase in patients who received a biventricular upgrade; however, Tp-e did increase in patients with preimplantation left bundle branch block or narrow QRS. Postimplantation Tp-e was the only independent predictor of appropriate ICD therapy.


Subject(s)
Defibrillators, Implantable , Electrocardiography , Signal Processing, Computer-Assisted , Ventricular Fibrillation/therapy , Adult , Aged , Bundle-Branch Block/therapy , Cardiac Output, Low/therapy , Death, Sudden, Cardiac/prevention & control , Female , Follow-Up Studies , Heart Failure/therapy , Hemodynamics/physiology , Humans , Kaplan-Meier Estimate , Male , Middle Aged , Prognosis , Recurrence , Retrospective Studies , Treatment Outcome , Ventricular Dysfunction, Left/therapy
19.
Heart Rhythm ; 4(10): 1300-5, 2007 Oct.
Article in English | MEDLINE | ID: mdl-17905335

ABSTRACT

BACKGROUND: Many patients with appropriate indications fail to respond to cardiac resynchronization therapy (CRT). OBJECTIVE: The purpose of our study was to determine the relationship between CRT response and preimplantation apical wall motion abnormality. METHODS: We analyzed data from 83 patients with ischemic cardiomyopathy who underwent CRT. All patients had New York Heart Association class III or IV symptoms despite maximal medical therapy, left ventricular ejection fraction (LVEF) < or =35%, and QRS duration > or =130 ms or <130 ms with left ventricular dyssynchrony. CRT responders at 6 months were defined as surviving patients with: (1) no hospitalization for heart failure, and (2) improvement of New York Heart Association classification. Patients underwent echocardiography before and 6 months after implantation to assess changes in regional wall motion and LVEF. RESULTS: At baseline, CRT responders (n = 39) and nonresponders (n = 44) had similar LVEF (22.9% +/- 6.9% vs 23.1% +/- 8.3%), QRS duration (159 +/- 43 ms vs 159 +/- 36 ms), and medical treatment. CRT nonresponders had a higher prevalence of preimplantation apical wall motion abnormality (68% vs 33%, P = .003). Patients with baseline apical wall motion abnormalities (n = 43) were less likely than others (n = 40) to show improvement in wall motion at 6 months (30% vs 81%, P < .001) or clinical response to CRT (31% vs 64%, P = .003). CONCLUSION: The presence of a preimplantation apical wall motion abnormality was associated with a lower rate of CRT response in patients with ischemic cardiomyopathy.


Subject(s)
Defibrillators, Implantable , Electrocardiography , Myocardial Ischemia/therapy , Ventricular Dysfunction, Left/therapy , Aged , Aged, 80 and over , Cardiac Output, Low/diagnostic imaging , Cardiac Output, Low/physiopathology , Cardiac Output, Low/therapy , Echocardiography , Female , Follow-Up Studies , Heart Failure/diagnostic imaging , Heart Failure/physiopathology , Heart Failure/therapy , Humans , Long QT Syndrome/diagnostic imaging , Long QT Syndrome/physiopathology , Long QT Syndrome/therapy , Male , Middle Aged , Myocardial Ischemia/diagnostic imaging , Myocardial Ischemia/physiopathology , Retrospective Studies , Treatment Failure , Ventricular Dysfunction, Left/diagnostic imaging , Ventricular Dysfunction, Left/physiopathology , Ventricular Function, Left/physiology
20.
J Am Coll Cardiol ; 50(14): 1324-31, 2007 Oct 02.
Article in English | MEDLINE | ID: mdl-17903630

ABSTRACT

OBJECTIVES: This study sought to characterize left atrial (LA) sinus rhythm electrogram (EGM) patterns and their relationship to parasympathetic responses during atrial fibrillation (AF) ablation. BACKGROUND: The mechanistic basis of fractionated LA EGMs in patients with paroxysmal AF is not well understood. METHODS: We analyzed 1,662 LA ablation sites from 30 patients who underwent catheter ablation for paroxysmal AF. Pre-ablation EGM characteristics (number of deflections, amplitude, and duration) were measured in sinus rhythm. Parasympathetic responses during radiofrequency application (increase of atrial-His interval by > or =10 ms or decrease of sinus rate by > or =20%) were assessed at all sites. We also prospectively studied the effect of adenosine, a pharmacological agent mimicking acetylcholine signaling in myocytes, on LA EGMs. Finally, we performed mathematical simulations of atrial tissue to delineate possible mechanisms of fractionated EGMs in sinus rhythm. RESULTS: A specific pattern of pre-ablation sinus rhythm EGM (deflections > or =4, amplitude > or =0.7 mV, and duration > or =40 ms) was strongly associated with parasympathetic responses (sensitivity 72%, specificity 91%). The sites associated with these responses were found to be located mainly in the posterior wall of the LA. Adenosine administration and mathematical simulation of the effect of acetylcholine were able to reproduce a similar EGM pattern. CONCLUSIONS: Parasympathetic activation during AF ablation is associated with the presence of pre-ablation high-amplitude fractionated EGMs in sinus rhythm. Local acetylcholine release could potentially explain this phenomenon.


Subject(s)
Atrial Fibrillation/diagnosis , Atrial Fibrillation/physiopathology , Electrocardiography/methods , Parasympathetic Nervous System , Sinoatrial Node/innervation , Adenosine/therapeutic use , Anti-Arrhythmia Agents/therapeutic use , Atrial Fibrillation/drug therapy , Atrial Fibrillation/surgery , Catheter Ablation , Female , Humans , Male , Middle Aged , Models, Biological , Retrospective Studies , Sensitivity and Specificity , Sinoatrial Node/drug effects
SELECTION OF CITATIONS