Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
J Electron Microsc Tech ; 15(4): 377-82, 1990 Aug.
Article in English | MEDLINE | ID: mdl-2391563

ABSTRACT

Single crystal MgO specimens having low load Vickers indentations were thinned in an ion milling machine employing a single ion gun, and their characteristics were investigated with optical microscopy and high voltage electron microscopy (HVEM). It was found that the state of cleanliness of the specimen chamber of the ion milling machine had a very marked influence on the quality of the thinned specimens. If the specimen chamber was not well cleaned before ion milling a fresh specimen, the latter tended to show amorphisation due to the deposition on the specimen of the debris left in the chamber from the previously ion-milled specimens. Such observations were made from MgO specimens ion milled in several different types of commercial ion milling machine employing a single gun. It is proposed that to obtain good-quality ion milled TEM specimens, it is important to clean the specimen chamber thoroughly prior to milling.


Subject(s)
Magnesium Oxide , Microscopy, Electron/methods , Specimen Handling , Surface Properties
SELECTION OF CITATIONS
SEARCH DETAIL
...