Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
Radiat Prot Dosimetry ; 187(4): 451-460, 2019 Dec 31.
Article in English | MEDLINE | ID: mdl-31650160

ABSTRACT

This study characterises and evaluates an Al2O3:C-based optically stimulated luminescent dosemeter (OSLD) system, commercially known as the nanoDot™ dosemeter and the InLight® microStar reader, for personal and in vivo dose measurements in diagnostic radiology. The system characteristics, such as dose linearity, reader accuracy, reproducibility, batch homogeneity, energy dependence and signal stability, were explored. The suitability of the nanoDot™ dosemeters was evaluated by measuring the depth dose curve, in vivo dose measurement and image perturbation. The nanoDot™ dosemeters were observed to produce a linear dose with ±2.8% coefficient variation. Significant batch inhomogeneity (8.3%) was observed. A slight energy dependence (±6.1%) was observed between 60 and 140 kVp. The InLight® microStar reader demonstrated good accuracy and a reproducibility of ±2%. The depth dose curve measured using nanoDot™ dosemeters showed slightly lower responses than Monte Carlo simulation results. The total uncertainty for a single dose measurement using this system was 11%, but it could be reduced to 9.2% when energy dependence correction was applied.


Subject(s)
Monte Carlo Method , Optically Stimulated Luminescence Dosimetry/instrumentation , Optically Stimulated Luminescence Dosimetry/methods , Radiation Dosimeters/standards , Radiography/methods , Calibration , Computer Simulation , Humans , Radiation Dosage , Reproducibility of Results
SELECTION OF CITATIONS
SEARCH DETAIL
...