1.
Radiol Med ; 126(7): 979-988, 2021 Jul.
Article in English | MEDLINE | ID: mdl-33900527

ABSTRACT

OBJECTIVE: Accurate calculation of set-up margin is a prerequisite to arrive at the most optimal clinical to planning target volume margin. The aim of this study was to evaluate the compatibility of different on-board and in-room stereoscopic imaging modalities by calculating the set-up margins (SM) in stereotactic body radiotherapy technique accounting and unaccounting for rotational positional errors (PE). Further, we calculated separate SMs one based on residual positional errors and another based on residual + intrafraction positional errors from the imaging data obtained in a dual imaging environment. MATERIALS AND METHODS: A total of 22 lung cancer patients were included in this study. For primary image guidance, four-dimensional cone beam computed tomography (4-D CBCT) was used and stereoscopic ExacTrac was used as the auxiliary imaging. Following table position correction (TPC) based on the initial 4-D CBCT, another 4-D CBCT (post-TPC) and a pair of stereoscopic ExacTrac images were obtained. Further, during the treatment delivery, a series of ExacTrac images were acquired to identify the intrafraction PE. If a, b and c were the observed translational shifts in lateral (x-axis), longitudinal (y-axis) and vertical direction (z-axis) and α, ß and γ were the rotational shifts in radians about the same axes, respectively, then the resultant translational vectors (A, B and C) were calculated on the basis of translational and rotational values. Set-up margins were calculated using residual errors post-TPC only and also using intrafraction positional errors in addition to the residual errors. RESULTS: Residual and residual + intrafraction SM were calculated from a dataset of 82 CBCTs and 189 ExacTrac imaging sessions. CBCT-based mean ± SD shifts in translational and rotational directions were 0.3 ± 1.8 mm, 0.1 ± 1.8 mm, - 0.4 ± 1.6 mm, 0.1 ± 0.4°, 0.0 ± 1.0° and 0.3 ± 0.7°, respectively, and for ExacTrac - 0.1 ± 1.8 mm, 0.2 ± 2.4 mm, - 0.6 ± 1.8 mm, 0.1 ± 1.2°, - 0.2 ± 1.3° and - 0.1 ± 0.6°, respectively. Residual SM without considering the rotational correction in x, y and z directions were 5.0 mm, 4.5 mm and 4.4 mm; rotation-corrected SM were 4.4 mm, 4.0 mm and 5.5 mm, respectively. Residual plus intrafraction SM were 5.5 mm, 6.6 mm and 6.2 mm without considering the rotational corrections, whereas they were 5.0 mm, 6.3 mm and 6.2 mm with rotational errors accounted for. CONCLUSION: Accurate calculation of set-up margin is required to find the clinical to planning target volume margin. Primary and auxiliary imaging margins fall in the range of 4.0 to 5.5 mm and 5.0 to 7.0 mm, respectively, indicating a higher SM for X-ray-based planar imaging techniques over three-dimensional cone beam images. This study established the degree of mutual compatibility between two different kinds of widely used set-up imaging modalities, on-board CBCT and in-room stereoscopic imaging ExacTrac. It also describes the technique to calculate the residual and residual plus intrafraction SM and its variation in a dual imaging environment accounting for rotational PE in stereotactic body radiotherapy of lung.


Subject(s)
Cone-Beam Computed Tomography/methods , Four-Dimensional Computed Tomography/methods , Lung Neoplasms/radiotherapy , Lung/diagnostic imaging , Radiotherapy Planning, Computer-Assisted/methods , Radiotherapy, Image-Guided/methods , Adult , Aged , Female , Humans , Lung Neoplasms/diagnosis , Male , Middle Aged
2.
Med Phys ; 46(11): 4749-4754, 2019 Nov.
Article in English | MEDLINE | ID: mdl-31495931

ABSTRACT

PURPOSE: The aim of this study is to calculate setup margin based on six-dimensional (6D) corrected residual positional errors from kV cone beam computed tomography (CBCT) and from intrafraction projection kV imaging in coplanar and in noncoplanar couch positions in stereotactic radiotherapy. METHODS: Six dimensional positional corrections were carried out before patient treatments, using a robotic couch and CBCT matching. A CBCT and stereoscopic ExacTrac image were acquired post-table position correction. Further, a series of intrafraction ExacTrac images were obtained for the variable couch position. Translational and rotational errors were identified as lateral (X), longitudinal (Y), vertical (Z); roll (Ɵ°), pitch (Φ°) and yaw (Ψ°). A total of 699 intrafraction image sets (361 coplanar and 338 noncoplanar) for 51 SRS/SRT patients were analysed. Rotational errors were corrected in terms of translational coordinates. Residual set-up margins were calculated from CBCT shifts. ExacTrac shifts give residual + intrafraction setup margins as a function of coplanar and noncoplanar couch positions. RESULTS: The average residual positional error obtained from CBCT in X, Y, Z, Ɵ, Φ, Ψ were 0.1 ± 0.4 mm, 0.0 ± 0.6 mm, 0.0 ± 0.5 mm, 0.2 ± 0.8°, 0.1 ± 0.6° and -0.1 ± 0.7° respectively. For ExacTrac, the shits were -0.5 ± 0.9 mm, -0.0 ± 1mm, -0.6 ± 1.0mm, 0.4 ± 0.9°, -0.2 ± 0.6°, and -0.0 ± 0.8°. CBCT calculated linear setup margins in X, Y, Z direction were 0.5, 1.2, and 1 mm respectively. ExacTrac yielded coplanar and noncoplanar linear setup margins were 1.2, 1.3, 1.5, 1.4, 1.5, and 2.1 mm respectively. CONCLUSION: CBCT-based gross residual set-up margin is equal to 1 mm. ExacTrac calculated residual plus intrafraction setup margin falls within a 2 mm range; attributed to intrafraction patient movement, table position inaccuracies, and poor image fusion in noncoplanar geometry. There could be variations in the required additional margin between centers and between machines, which require further studies.


Subject(s)
Dose Fractionation, Radiation , Radiosurgery/methods , Radiotherapy Setup Errors , Rotation , Cone-Beam Computed Tomography , Humans , Neoplasms/diagnostic imaging , Neoplasms/radiotherapy