J Appl Clin Med Phys ; 4(4): 307-14, 2003.
Article in English | MEDLINE | ID: mdl-14604420

ABSTRACT

The GafChromic XR Type R film is a relatively new product for recording high radiation dose in interventional radiological procedures. Means of measuring the film response were studied in this investigation. Two handheld reflective-type densitometers of different models were compared in the range of 0-8 Gy. They were found to be in excellent agreement. Five reflective flatbed scanners of different models were compared by a simple preliminary test. Their widely differed performances suggest the need of testing a scanner before using it for dosimetry measurement. A selected scanner was further tested for its ability to measure radiation in the range of 0-8 Gy and for the development of a scanning protocol. This experiment suggested the inclusion of a calibration pattern with known exposures and a black reference step in the scanning of a film in RGB mode. Then the red component of this image should be used for dosimetry computation. This method was compared to the use of a red acetate filter. The latter was demonstrated to be a possible alternative for measurement below 5 Gy and when there is no software ability to split an image into color components.


Subject(s)
Absorptiometry, Photon/instrumentation , Absorptiometry, Photon/methods , X-Ray Film/standards , Absorptiometry, Photon/standards , Calibration , Film Dosimetry/instrumentation , Film Dosimetry/methods , Film Dosimetry/standards , Radiation Dosage