1.
J Bone Joint Surg Br ; 84(8): 1199-204, 2002 Nov.
Article in English | MEDLINE | ID: mdl-12463671

ABSTRACT

We have compared the interface morphology at the stem-cement interface of standard Charnley stems with a satin finish (Ra = 0.75 microm) with identical stems which had been grit-blasted over their proximal third (Ra = 5.3 microm) to promote a proximal bond. The stems were cemented into cadaver femora using conventional contemporary cementing techniques. After transverse sectioning, we determined the percentage of the perimeter of the stem which had a gap at the interface. There were substantial gaps (mean 31.4 +/- 17.1%) at the stem-cement interface in the grit-blasted region. This fraction was significantly (paired t-test, p = 0.0054) higher than that found around the contralateral satin-finished stems (mean 7.7 +/- 11.7%). Although studies of isolated metal-cement interfaces have shown that the bond strength can increase with surface roughness it cannot be assumed that this effect will be observed under clinical conditions.


Subject(s)
Bone Cements , Femur , Hip Prosthesis , Analysis of Variance , Cadaver , Humans , Microscopy, Electron , Prosthesis Failure , Surface Properties