Ann Biol Clin (Paris) ; 77(4): 447-452, 2019 08 01.
Article in French | MEDLINE | ID: mdl-31339492

ABSTRACT

The complement system is composed of a set of plasma or membrane proteins. Complement protein deficiencies can be inherited or acquired, through the presence of autoantibodies or through consumption. We evaluated the analytical performance of the Optilite® analyser for the determination of the C3 and C4 levels and for the evaluation of the total complement activity. The intra- and inter-series CVs were evaluated and have showed satisfactory results, the concordances with analysers currently used in the laboratory (BNII® and BCT®, Siemens) are very good, as is the agreement between the serum and plasma samples. We also determined the reference values for the different parameters tested in view of a routine use of Optilite® analyser in the laboratory.


Subject(s)
Complement C3/analysis , Complement C4/analysis , Complement System Proteins/analysis , Hematologic Tests/instrumentation , Adult , Artifacts , Automation, Laboratory/instrumentation , Automation, Laboratory/standards , Blood Chemical Analysis/instrumentation , Blood Chemical Analysis/methods , Blood Chemical Analysis/standards , Blood Coagulation/physiology , Complement C3/metabolism , Complement C4/metabolism , Complement System Proteins/metabolism , Equipment Contamination , Hematologic Tests/methods , Hematologic Tests/standards , Humans , Reference Standards , Reproducibility of Results