Eur J Cardiovasc Nurs ; 22(8): 773-779, 2023 Dec 14.
Article in English | MEDLINE | ID: mdl-36346102

ABSTRACT

AIMS: Implantable cardiac monitors (ICMs) can generate false-positive (FP) alerts. Although these devices have an extended programmability, there are no recommendations on their optimization to reduce not-relevant activations.We tested a strategic programming optimization guide based on the type of FP and investigated the safety and feasibility of the nurse-led insertion of ICMs with a long-sensing vector. METHODS AND RESULTS: Consecutive patients implanted by trained nurses with long-sensing vector ICM were enrolled in a 1-month observational stage (Phase A). Patients who had ≥10 FP episodes underwent ICM reprogramming based on the predefined guide and were followed for an additional month (Phase B). A total of 78 patients had successful ICM insertion by nurses with a mean R wave amplitude of 0.96 ± 0.43 mV and an 86% P wave visibility. Only one patient reported a significant device-related issue, and nurse-delivered ICM was generally well accepted by the patients. During Phase A, 11 patients (14%) generated most of FP (3,627/3,849; 94%) and underwent ICM reprogramming. In the following month (Phase B), five patients (45%) were free from FP and six (55%) transmitted 57 FP alerts (98% reduction compared with Phase A). The median number of FP per patient was significantly reduced after reprogramming [195 (interquartile range, 50-311) vs. one (0-10), P = 0.0002]. CONCLUSION: A strategic reprogramming of ICM in those patients with a high FP alert burden reduces the volume of erroneous activations with potential benefits for the remote monitoring service. No concerns were raised regarding nurse-led insertion of ICMs with a long-sensing vector.


Subject(s)
Atrial Fibrillation , Electrocardiography, Ambulatory , Humans , Arrhythmias, Cardiac , Electrocardiography , Electrocardiography, Ambulatory/methods , Nurse's Role , Prostheses and Implants , Feasibility Studies
2.
Int J Cardiol ; 334: 49-54, 2021 Jul 01.
Article in English | MEDLINE | ID: mdl-33930512

ABSTRACT

BACKGROUND: The remote follow-up of pacemakers and implantable cardiac defibrillators (ICDs) usually includes scheduled checks and alert transmissions. However, this results in a high volume of remote data reviews to be managed. We measured the relative contribution of scheduled and alert transmissions to the detection of relevant conditions, and the workload generated by their management. METHODS: At our center, the frequency of remote scheduled transmissions is 4/year. Moreover, all system-integrity and clinical alerts are turned on for wireless notification. We calculated the number of transmissions received from January to December 2020, and identified transmissions that necessitated in-hospital access for further assessment and transmissions that required clinical discussion with the physician. For all alert transmissions, we identified whether the alert was clinically meaningful (i.e. center was not previously aware of the condition and no action had yet been taken to treat it). RESULTS: Of 8545 transmissions received from 1697 pacemakers and ICDs, 5766 (67%) were scheduled and 2779 (33%) were alert transmissions received from 764 patients (45%); 499 (9%) scheduled transmissions required clinical discussion with the physician, but only 2 of these necessitated in-hospital visits for further assessment. Of the alert transmissions, 664 (24%) required clinical discussion, and 75 (3%) necessitated in-hospital visits. The proportion of alerts judged clinically meaningful was 7%. CONCLUSION: Scheduled transmissions generate 67% of remote data reviews for pacemakers and ICDs, but their ability to detect clinically relevant events is very low. A strategy that relies exclusively on alert transmissions could ensure continuity of patient monitoring while reducing the workload at the center.


Subject(s)
Defibrillators, Implantable , Electronics , Follow-Up Studies , Humans , Monitoring, Physiologic , Workload