Perfusion ; 17(6): 447-50, 2002 Nov.
Article in English | MEDLINE | ID: mdl-12470036

ABSTRACT

We present here a technique to replace a failed oxygenator by inserting a second oxygenator in parallel (PRONTO) within the cardiopulmonary bypass (CPB) circuit. Oxygenator failure is a potential hazard that may result in patient injury or death. Although failures are rare, safety surveys conducted over the last 25 years suggest that the incidence of oxygenator failures is on the rise. This emergency procedure may be easily applied to any standard CPB circuit with a few minor alterations. The technique is simple; it can be carried out rapidly. An important advantage of this technique is that it may be executed without interrupting blood flow to the patient, which may reduce the incidence of patient injury or death.


Subject(s)
Cardiopulmonary Bypass , Emergency Medical Services/methods , Oxygenators/adverse effects , Cardiopulmonary Bypass/instrumentation , Equipment Failure , Humans
2.
Perfusion ; 15(2): 129-35, 2000 Mar.
Article in English | MEDLINE | ID: mdl-10789567

ABSTRACT

Low hematocrit (Hct < 20) during cardiopulmonary bypass (CPB) is associated with higher mortality and other adverse outcomes. More frequently, low Hct is encountered in patients with small body size and women patients. This prompted us to take an aggressive approach in our care of these patients, involving a strategy for predicting patients at risk of low Hct, with the aid of an electronic worksheet that accurately predicts CPB Hct, and two prevention strategies: use of a low-prime CPB circuit (LP) for all adult patients with a body surface area (BSA) < 1.7 m(2) and use of autologous circuit priming (AP), in addition to the low-circuit volume in some patients. The two cohorts of patients in whom these techniques were employed were compared to a group matched for body size where our standard adult circuit (STD) was used. There were 233 patients in the standard group, 139 in the LP group, and 68 in the LP/AP group. The CPB circuit prime volume was 1,710 ml for the STD group and 1,110 ml for the LP group. Use of autologous priming techniques further reduced the prime volume by 545 +/- 139 ml. The incidence of low Hct (<20%) during CPB was thus reduced from 70% to 15% (p = 0.001) when using both techniques together without increasing red blood cell (RBC) transfusions. These changes in perfusion management resulted in a reduction in the incidence of renal complications (STD = 9.4%, LP = 6.5% (ns) and LP/AP = 0%,


Subject(s)
Blood Transfusion, Autologous , Body Constitution , Cardiopulmonary Bypass/instrumentation , Hemodilution/methods , Adult , Aged , Algorithms , Blood Volume , Body Height , Body Weight , Equipment Design , Evaluation Studies as Topic , Female , Hematocrit , Humans , Kidney Function Tests , Male