ASAIO J ; 52(3): 248-56, 2006.
Article in English | MEDLINE | ID: mdl-16760712

ABSTRACT

This study examined the effect of artificial lung compliance (C) on pulmonary system (PS) impedance and right ventricular function during in-series attachment of the MC3 Biolung in adult sheep. Compliances, C, of 0-20 ml/mm Hg were tested at the Biolung inlet. Results indicate the PS 0 harmonic input impedance modulus was not affected by C. The PS first harmonic input impedance modulus (Z1) was 10.9 +/- 3.2 mm Hg/(l/min) at C = 0 ml/mm Hg and minimized to 2.41 +/- 0.79 mm Hg/(l/min) at C > or = 0.5 ml/mm Hg. Cardiac output was 58% +/- 10% of its pre-Biolung attachment, baseline value at C = 0 ml/mm Hg and was maximized to an average of 75% +/- 11% at C > or = 0.5 ml/mm Hg. The left ventricular lateral-to-anteroposterior axis length ratio, which decreases with leftward septal shift, increased with C from 0.52 +/- 0.12 at C = 0 ml/mm Hg to 0.76 +/- 0.06 at C = 5 ml/mm Hg (p < 0.05), but decreased slightly with C at C > 5 ml/mm Hg. Therefore, the ideal C for right ventricular function is at least 0.5 ml/mm Hg and may be as high as 5 ml/mm Hg to minimize septal shift.


Subject(s)
Artificial Organs , Hemodynamics/physiology , Lung Compliance , Lung/physiology , Pulmonary Circulation , Animals , Biomedical Engineering , Electrocardiography , Male , Models, Cardiovascular , Prosthesis Design , Pulmonary Gas Exchange , Respiratory Mechanics , Sheep , Ventricular Function, Right/physiology