1.
Phys Med ; 42: 292-297, 2017 Oct.
Article in English | MEDLINE | ID: mdl-28736285

ABSTRACT

Simulations of planar whole body acquisitions in therapeutic procedures are often extensively time-consuming and therefore rarely used. However, optimising tools and variance reduction techniques can be employed to overcome this problem. In this paper, a variety of features available in GATE are explored and their capabilities to reduce simulation time are evaluated. For this purpose, the male XCAT phantom was used as a virtual patient with 177Lu-DOTATATE pharmacokinetic for whole body planar acquisition simulations in a Siemens Symbia T2 model. Activity distribution was divided into 8 compartments that were simulated separately. GATE optimization techniques included reducing the amount of time spent in both voxel and detector tracking. Some acceleration techniques led to a decrease of CPU-time by a factor of 167, while image statistics were kept constant. In that context, the simulation of therapeutic procedure imaging would still require 46days on a single CPU, but this could be reduced to hours on a dedicated cluster.


Subject(s)
Computer Simulation , Octreotide/analogs & derivatives , Organometallic Compounds , Phantoms, Imaging , Radionuclide Imaging/methods , Radiopharmaceuticals , Whole Body Imaging/methods , Humans , Kinetics , Male , Monte Carlo Method , Radionuclide Imaging/instrumentation , Whole Body Imaging/instrumentation
SELECTION OF CITATIONS