Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
Ann Clin Biochem ; 50(Pt 5): 416-20, 2013 Sep.
Article in English | MEDLINE | ID: mdl-23828944

ABSTRACT

BACKGROUND: Macrocomplexes can be the cause of elevated serum hormone concentrations and may cause diagnostic confusion. This is well recognized for prolactin and commonly screened for using polyethylene glycol (PEG) precipitation. The phenomenon and a suitable screening method is less familiar with respect to thyroid-stimulating hormone (TSH). METHOD: Samples sent to the laboratory for routine analysis of thyroid function and found to have a TSH >10 mU/L were evaluated to determine the prevalence of macro-TSH in the Roche Elecsys assay, using PEG precipitation with confirmation by gel filtration chromatography. RESULTS: Of 495 samples tested, 3 (0.6%) were found to have macro-TSH. From the distribution of recoveries, a cut-off <25% was determined for identifying samples requiring further investigation for the presence of macro-TSH. CONCLUSION: The prevalence of elevated TSH due to macro-TSH was found to be 0.6%. Laboratories should be aware of this cause of assay interference.


Subject(s)
Antigen-Antibody Complex/blood , Congenital Hypothyroidism/blood , Immunoglobulin G/blood , Thyrotropin/blood , Antigen-Antibody Complex/chemistry , Chromatography, Gel , Congenital Hypothyroidism/diagnosis , Congenital Hypothyroidism/immunology , False Positive Reactions , Humans , Immunoglobulin G/chemistry , Polyethylene Glycols/chemistry , Protein Binding , Thyroid Function Tests , Thyrotropin/chemistry
SELECTION OF CITATIONS
SEARCH DETAIL
...