Microsc Microanal ; 26(1): 63-75, 2020 Feb.
Article in English | MEDLINE | ID: mdl-31858928

ABSTRACT

When selecting a method for determining modal mineralogy and elemental composition of geological samples (e.g., meteorites), a number of factors should be considered, includingthe number of objects or the area to be analyzed; the scale of expected chemical variation; instrument time restrictions; and post-processing time. This study presents a method that minimizes acquisition time while maintaining the ability to distinguish minerals based on combinations of intensities of electron probe micro-analyzer-generated X-ray element maps. While some other methods yield similar outcomes, this method's post-processing utilizes standard parameterized, X-ray intensity "map math" in an algorithm that is adaptable and requires minimal supervision once implemented. This study's minimized supervision in the post-processing of X-ray intensity maps decreases analysis time and its adaptability increases the number of potential applications. The method also facilitates calibration of the exact locations of analysis using laser ablation methods. While the method described here has advantages, the choice of method always depends on the question being asked.