Radiat Prot Dosimetry ; 169(1-4): 319-24, 2016 Jun.
Article in English | MEDLINE | ID: mdl-26538617

ABSTRACT

As opposed to the standard detective quantum efficiency (DQE), effective DQE (eDQE) is a figure of merit that allows comparing the performances of imaging systems in the presence of scatter rejection devices. The geometry of the EOS™ slot-scanning system is such that the detector is self-collimated and rejects scattered radiation. In this study, the EOS system was characterised using the eDQE in imaging conditions similar to those used in clinical practice: with phantoms of different widths placed in the X-ray beam, for various incident air kerma and tube voltages corresponding to the phantom thickness. Scatter fractions in EOS images were extremely low, around 2 % for all configurations. Maximum eDQE values spanned 9-14.8 % for a large range of air kerma at the detector plane from 0.01 to 1.34 µGy. These figures were obtained with non-optimised EOS setting but still over-performed most of the maximum eDQEs recently assessed for various computed radiology and digital radiology systems with antiscatter grids.


Subject(s)
Imaging, Three-Dimensional/instrumentation , Radiographic Image Enhancement/instrumentation , Radiographic Image Interpretation, Computer-Assisted/instrumentation , Tomography, X-Ray Computed/instrumentation , Equipment Design , Equipment Failure Analysis , Phantoms, Imaging , Radiation Dosage , Reproducibility of Results , Sensitivity and Specificity , Transducers