1.
Article in English | MEDLINE | ID: mdl-39230796

ABSTRACT

PURPOSE: To address the clinical need for totally implantable mechanical circulatory support devices, Bionet Sonar is developing a novel Ultrasonic Transcutaneous Energy Transmission (UTET) system that is designed to eliminate external power and/or data communication drivelines. METHODS: UTET systems were designed, fabricated, and pre-clinically tested using a non-clinical HeartWare HVAD in static and dynamic mock flow loop and acute animal models over a range of pump speeds (1800, 2400, 3000 RPM) and tissue analogue thicknesses (5, 10, 15 mm). RESULTS: The prototypes demonstrated feasibility as evidenced by meeting/exceeding function, operation, and performance metrics with no system failures, including achieving receiver (harvested) power exceeding HVAD power requirements and data communication rates of 10kB/s and pump speed control (> 95% sensitivity and specificity) for all experimental test conditions, and within healthy tissue temperature range with no acute tissue damage. CONCLUSION: During early-stage development and testing, engineering challenges for UTET size reduction and stable and safe operation were identified, with solutions and plans to address the limitations in future design iterations also presented.

2.
Neuromodulation ; 2024 May 30.
Article in English | MEDLINE | ID: mdl-38819342

ABSTRACT

OBJECTIVES: This study aimed to indicate the feasibility of a prototype electrical neuromodulation system using a closed-loop energy-efficient ultrasound-based mechanism for communication, data transmission, and recharging. MATERIALS AND METHODS: Closed-loop deep brain stimulation (DBS) prototypes were designed and fabricated with ultrasonic wideband (UsWB) communication technology and miniaturized custom electronics. Two devices were implanted short term in anesthetized Göttingen minipigs (N = 2). Targeting was performed using preoperative magnetic resonance imaging, and locations were confirmed postoperatively by computerized tomography. DBS systems were tested over a wide range of stimulation settings to mimic minimal, typical, and/or aggressive clinical settings, and evaluated for their ability to transmit data through scalp tissue and to recharge the DBS system using UsWB. RESULTS: Stimulation, communication, reprogramming, and recharging protocols were successfully achieved in both subjects for amplitude (1V-6V), frequency (50-250 Hz), and pulse width (60-200 µs) settings and maintained for ≥six hours. The precision of pulse settings was verified with <5% error. Communication rates of 64 kbit/s with an error rate of 0.05% were shown, with no meaningful throughput degradation observed. Time to recharge to 80% capacity was <9 minutes. Two DBS systems also were implanted in the second test animal, and independent bilateral stimulation was successfully shown. CONCLUSIONS: The system performed at clinically relevant implant depths and settings. Independent bilateral stimulation for the duration of the study with a 4F energy storage and full rapid recharge were achieved. Continuous function extrapolates to six days of continuous stimulation in future design iterations implementing application specific integrated circuit level efficiency and 15F storage capacitance. UsWB increases energy efficiency, reducing storage requirements and thereby enabling device miniaturization. The device can enable intelligent closed-loop stimulation, remote system monitoring, and optimization and can serve as a power/data gateway to interconnect the intrabody network with the Internet of Medical Things.

SELECTION OF CITATIONS
SEARCH DETAIL