J Chromatogr A ; 1219: 195-200, 2012 Jan 06.
Article in English | MEDLINE | ID: mdl-22138227

ABSTRACT

A novel capillary electrophoresis (CE) method coupled with monolithic molecular imprinted polymer (MIP) fiber based solid phase microextraction (SPME) was developed for selective and sensitive determination of ephedrine (E) and pseudoephedrine (PE). With in situ polymerization in a silica capillary mold and E as template, the MIP fibers could be produced in batch reproducibly and each fiber was available for 50 extraction cycles without significant decrease in extraction ability. Using the MIP fiber under optimized extraction conditions, CE detection limits of E and PE were greatly lowered from 0.20 to 0.00096 µg/mL and 0.12 to 0.0011 µg/mL, respectively. Analysis of urine and serum samples by the MIP-SPME-CE method was also performed, with results indicating that E and PE could be selectively extracted. The recoveries and relative standard deviations (RSDs) for sample analysis were found in the range of 91-104% and 3.8-9.1%, respectively.


Subject(s)
Electrophoresis, Capillary/methods , Ephedrine/isolation & purification , Molecular Imprinting/methods , Pseudoephedrine/isolation & purification , Solid Phase Microextraction/methods , Acetic Acid , Ephedrine/blood , Ephedrine/urine , Humans , Limit of Detection , Methanol , Pseudoephedrine/blood , Pseudoephedrine/urine , Reproducibility of Results , Sodium Chloride , Solid Phase Microextraction/instrumentation , Time Factors