1.
Pacing Clin Electrophysiol ; 22(10): 1481-7, 1999 Oct.
Article in English | MEDLINE | ID: mdl-10588150

ABSTRACT

Ventricular fibrillation (VF) duration may be a factor in determining the defibrillation energy for successful defibrillation. Exponential biphasic waveforms have been shown to defibrillate with less energy than do monophasic waveforms when used for external defibrillation. However, it is unknown whether this advantage persists with longer VF duration. We tested the hypothesis that exponential biphasic waveforms have lower defibrillation energy as compared to exponential monophasic waveforms even with longer VF duration up to 1 minute. In a swine model of external defibrillation (n = 12, 35 +/- 6 kg), we determined the stored energy at 50% defibrillation success (E50) after both 10 seconds and 1 minute of VF duration. A single exponential monophasic (M) and two exponential biphasic (B1 and B2) waveforms were tested with the following characteristics: M (60 microF, 70% tilt), B1 (60/60 microF, 70% tilt/3 ms pulse width), and B2 (60/20 microF, 70% tilt/3 ms pulse width) where the ratio of the phase 2 leading edge voltage to that of phase 1 was 0.5 for B1 and 1.0 for B2. E50 was measured by a Bayesian technique with a total often defibrillation shocks in each waveform and VF duration randomly. The E50 (J) for M, B1, and B2 were 131 +/- 41, 57 +/- 18,* and 60 +/- 26* with 10 seconds of VF duration, respectively, and 114 +/- 62, 77 +/- 45,* and 72 +/- 53* with 1 minute of VF duration, respectively (*P < 0.05 vs M). There was no significant difference in the E50 between 10 seconds and 1 minute of VF durations for each waveform. We conclude that (1) the E50 does not significantly increase with lengthening VF durations up to 1 minute regardless of the shock waveform, and (2) external exponential biphasic shocks are more effective than monophasic waveforms even with longer VF durations.


Subject(s)
Electric Countershock/methods , Ventricular Fibrillation/physiopathology , Analysis of Variance , Animals , Heart/physiology , Heart/physiopathology , Swine , Treatment Outcome , Ventricular Fibrillation/therapy
2.
Pacing Clin Electrophysiol ; 22(6 Pt 1): 849-54, 1999 Jun.
Article in English | MEDLINE | ID: mdl-10392381

ABSTRACT

The optimal electrode configuration for endocardial defibrillation is still a matter of debate. Current data suggests that a two pathway configuration using the right ventricle (RV) as cathode and a common anode constituted by a superior vena cava (SVC) and a pectoral can (C) is the most effective combination. This may be related to the more uniform voltage gradient created by shocks delivered using this configuration. We hypothesized that more effective waveforms could be obtained by varying the distribution of the shock current between the two pathways of a three electrode endocardial defibrillation system. In 12 pigs, we compared the characteristics and the defibrillation efficacy of six biphasic waveforms discharged using either a two (RV-->C) or a three (RV-->SVC + C) electrode combination with the following configurations: Configuration 1 (W1): the RV apical coil was used as a cathode and the subcutaneous C as anode (RV-->C). Configuration 2 (W2): The RV was used as cathode and the combination of the atriocaval coil (SVC) and the subcutaneous C as anode (RV-->SVC + C). Configuration 3 (W3): The RV-->C was used for the first 25% of f+ and RV-->SVC + C for the remainder of the discharge including f 2 Configuration 4 (W4): The RV-->C was used for the first 50% of f+ and RV-->SVC + C for the remainder of the discharge including f 2 Configuration 5 (W5): The RV-->C was used for the first 75% of f+ and RV-->SVC + C for the remainder of the discharge including f 2. Configuration 6 (W6): The RV-->C was used for f+ and RV-->SVC + C for f 2. As an increasing fraction of the waveform was discharged using the RV-->SVC + C pathways, the impedance and the pulse width decreased while the tilt, the peak, and the average current significantly increased. The waveforms delivered using the RV-->SVC + C configuration for 100% or 75% of their duration had significantly lower stored energy DFT than the other waveform. Current distribution between three endocardial electrodes can be altered during the shock and generates waveforms with different characteristics. Shocks with 75% or more of the current flowing to the RV-->SVC + C required the lowest stored energy to defibrillate. This method of energy steering could be used to optimize current delivery in a three electrodes system.


Subject(s)
Defibrillators, Implantable , Electrocardiography , Animals , Electrocardiography/instrumentation , Electrodes , Endocardium/physiopathology , Equipment Design , Heart Ventricles/physiopathology , Signal Processing, Computer-Assisted/instrumentation , Swine
3.
Circulation ; 98(22): 2487-93, 1998 Dec 01.
Article in English | MEDLINE | ID: mdl-9832496

ABSTRACT

BACKGROUND: Biphasic waveforms have been reported to be more efficacious than monophasic waveforms for external defibrillation. This study examined the optimal phase-1 tilts and phase-2 leading-edge voltages with small capacitors (60 and 20 microF) for external defibrillation. We also assessed the ability of the "charge-burping" model to predict the optimal waveforms. METHODS AND RESULTS: Two groups of studies were performed. In group 1, 9 biphasic waveforms from a combination of 3 phase-1 tilt values (30%, 50%, and 70%) and 3 phase-2 leading-edge voltage values (0.5, 1.0, and 1.5 times the phase-1 leading-edge voltage, V1) were tested. Phase-2 pulse width was held constant at 3 ms in all waveforms. Two separate 60- microF capacitors were used in each phase. The energy value that would produce a 50% likelihood of successful defibrillation (E50) decreased with increasing phase-1 tilt and increased with increasing phase-2 leading-edge voltage except for the 30% phase-1 tilt waveforms. In group 2, 9 waveforms were identical to the waveforms in group 1, except for a 20- microF capacitor for phase 2. E50 decreased with increasing phase-1 tilt. Phase-2 leading-edge voltage of 1.0 to 1.5 V1 appeared to minimize E50 for phase-1 tilt of 50% and 70% but worsened E50 for phase-1 tilt of 30%. There was a significant correlation between E50 and residual membrane voltage at the end of phase 2, as calculated by the charge-burping model in both groups (group 1, R2=0.47, P<0.001; group 2, R2=0.42, P<0.001). CONCLUSIONS: The waveforms with 70% phase-1 tilt were more efficacious than those with 30% and 50%. The relationship of phase-2 leading-edge voltage to defibrillation efficacy depended on phase-2 capacitance. The charge-burping model predicted the optimal external biphasic waveform.


Subject(s)
Electric Countershock/instrumentation , Animals , Electric Conductivity , Models, Cardiovascular , Swine
4.
Pacing Clin Electrophysiol ; 21(6): 1216-24, 1998 Jun.
Article in English | MEDLINE | ID: mdl-9633063

ABSTRACT

The impedance of defibrillation pathways is an important determinant of ventricular defibrillation efficacy. The hypothesis in this study was that the respiration phase (end-inspiration versus end-expiration) may alter impedance and/or defibrillation efficacy in a "hot can" electrode system. Defibrillation threshold (DFT) parameters were evaluated at end-expiration and at end-inspiration phases in random order by a biphasic waveform in ten anesthetized pigs (body weight: 19.1 +/- 2.4 kg; heart weight: 97 +/- 10 g). Pigs were intubated with a cuffed endotracheal tube and ventilated through a Drager SAV respirator with tidal volume of 400-500 mL. A transvenous defibrillation lead (6 cm long, 6.5 Fr) was inserted into the right ventricular apex. A titanium can electrode (92-cm2 surface area) was placed in the left pectoral area. The right ventricular lead was the anode for the first phase and the cathode for the second phase. The DFT was determined by a "down-up down-up" protocol. Statistical analysis was performed with a Wilcoxon matched pair test. The median impedance at DFT for expiration and inspiration phases were 37.8 +/- 3.1 omega, and 39.3 +/- 3.6 omega, respectively (P = 0.02). The stored energy at DFT for expiration and inspiration phases were 5.7 +/- 1.9 J and 6.0 +/- 1.0 J, respectively (P = 0.594). Shocks delivered at end-inspiration exhibited a statistically significant increase in electrode impedance in a " hot can" electrode system. The finding that DFT energy was not significantly different at both respiration phases indicates that respiration phase does not significantly affect defibrillation energy requirements.


Subject(s)
Electric Countershock , Respiration , Ventricular Fibrillation/therapy , Animals , Electric Countershock/instrumentation , Electric Countershock/methods , Electric Impedance , Heart Conduction System/physiopathology , Intubation, Intratracheal , Swine , Ventricular Fibrillation/physiopathology
5.
J Cardiovasc Electrophysiol ; 8(6): 649-57, 1997 Jun.
Article in English | MEDLINE | ID: mdl-9209966

ABSTRACT

INTRODUCTION: A biphasic defibrillation waveform can achieve a large second phase leading-edge voltage by a "parallel-series" switching system. Recently, such a system using two 30-microF capacitances demonstrated better defibrillation threshold than standard waveforms available in current implantable devices. However, the optimized tilt of such a "parallel-series" system had not been defined. METHODS AND RESULTS: Defibrillation thresholds were evaluated for five different biphasic "parallel-series" waveforms (60/15 microF) and a biphasic "parallel-parallel" waveform (60/60 microF) in 12 anesthetized pigs. The five "parallel-series" waveforms had first phase tilts of 40%, 50%, 60%, 70%, and 80% with second phase pulse width of 3 msec. The "parallel-parallel" waveform had first phase tilt of 50% with second phase pulse width of 3 msec. The defibrillation lead system comprised a left pectoral "hot can" electrode (cathode) and a right ventricular lead (anode). The stored energy at defibrillation threshold of the "parallel-series" waveform with first phase tilts of 40%, 50%, 60%, 70%, and 80% was 7.0 +/- 2.1, 6.1 +/- 2.8, 6.8 +/- 2.8, 7.2 +/- 2.9, and 8.4 +/- 3.1 J, respectively. The stored energy of the "parallel-series" waveform with a 50% first phase tilt was 16% less than the nonswitching "parallel-parallel" waveform (7.3 +/- 2.8 J, P = 0.006). CONCLUSIONS: A first phase tilt of 50% maximized defibrillation efficacy of biphasic waveforms implemented with a "parallel-series" switching system. This optimized "parallel-series" waveform was more efficient than the comparable "parallel-parallel" biphasic waveform having the same first phase capacitance and tilt.


Subject(s)
Defibrillators, Implantable , Electric Countershock/methods , Animals , Electric Conductivity , Swine
6.
J Cardiovasc Electrophysiol ; 8(5): 517-28, 1997 May.
Article in English | MEDLINE | ID: mdl-9160228

ABSTRACT

INTRODUCTION: A major limitation in a conventional truncated exponential waveform is the rapid drop in current that results in short duration of high current or longer duration with a lower average current. We hypothesized that increasing the first phase average current by boosting the decaying waveform prior to phase reversal may improve defibrillation efficacy. METHODS AND RESULTS: To better simulate a "rectangular" waveform during the first phase, a "sawtooth" defibrillation waveform was constructed using "parallel-series" switching of capacitances (each 30 microF) during the first phase. This permitted a boost in the voltage late in the first phase. This sawtooth biphasic waveform (sawtooth) was compared to two clinical waveforms: a 135-microF capacitance (control-1) and a 90-microF capacitance (control-2) waveform. Defibrillation threshold (DFT) parameters were evaluated in 13 anesthetized pig models using a system consisting of a transvenous right ventricular apex lead (anode) and a left pectoral "hot can" electrode (cathode) system. DFT was determined by a "down-up down-up" protocol. The stored energy for sawtooth, control-1, and control-2 was 10.5 +/- 2.8 J, 12.3 +/- 3.7 J*, and 12.2 +/- 2.8 J*, respectively (*P < or = 0.01 vs sawtooth). The average current of the first phase for sawtooth, control-1, and control-2 was 7.6 +/- 1.3 A, 4.7 +/- 0.9 A*, and 6.2 +/- 0.9 A*, respectively (*P = 0.0001 vs sawtooth). CONCLUSION: A sawtooth biphasic waveform utilizing a "parallel-series" switching system of smaller capacitors can improve defibrillation efficacy. A higher average current in the first phase generated by such a waveform may contribute to more efficient defibrillation by facilitating myocyte capture.


Subject(s)
Electric Countershock/methods , Animals , Electric Countershock/instrumentation , Electric Impedance , Electrocardiography , Electrodes , Models, Biological , Swine , Ventricular Dysfunction, Left/physiopathology , Ventricular Function
7.
Pacing Clin Electrophysiol ; 20(12 Pt 1): 2911-6, 1997 Dec.
Article in English | MEDLINE | ID: mdl-9455750

ABSTRACT

The polarity of a monophasic and biphasic shocks have been reported to influence DFTs in some studies. The purpose of this study was to evaluate the effect of the first phase polarity on the DFT of a biphasic shock utilizing a nonthoracotomy "hot can" electrode configuration which had a 90-microF capacitance. We tested the hypothesis that anodal first phase was more effective than cathodal ones for defibrillation using biphasic shocks in ten anesthetized pigs weighing 38.9 +/- 3.9 kg. The lead system consisted of a right ventricular catheter electrode with a surface area of 2.7 cm2 and a left pectoral "hot can" electrode with 92.9 cm2 surface area. DFT was determined using a repeated "down-up" technique. A shock was tested 10 seconds after initiation of ventricular fibrillation. The mean delivered energy at DFT was 11.2 +/- 1.7 J when using the right ventricular apex electrode as the cathode and 11.3 +/- 1.2 J (P = NS) when using it as the anode. The peak voltage at DFT was also not significantly different (529.0 +/- 41.3 and 531.8 +/- 28.6 V, respectively). We concluded that the first phase polarity of a biphasic shock used with a nonthroracotomy "hot can" electrode configuration did not affect DFT.


Subject(s)
Electric Countershock , Electrodes , Ventricular Fibrillation/therapy , Animals , Cardiac Catheterization , Disease Models, Animal , Electric Countershock/instrumentation , Electric Countershock/methods , Electrocardiography , Swine , Treatment Outcome , Ventricular Fibrillation/physiopathology
8.
Pacing Clin Electrophysiol ; 19(12 Pt 1): 2051-9, 1996 Dec.
Article in English | MEDLINE | ID: mdl-8994943

ABSTRACT

ICDs must sense R waves over a range of amplitudes without sensing P or T waves. Automatic threshold control (ATC) is an accepted sensing method for that task. ATC sensing levels are from 25%-75% of the electrogram (EGM) peak, decreasing with an exponential decay. A high sensing level for a time after peak detection may better allow ATC to pass over a T wave, while a lower sensing level thereafter may better allow ATC to sense the next R wave. An ATC was designed with two sensing levels and time constants (tau), using a 58% level (tau = 1.75 s) for 325 ms after peak detection switching to 33% (tau = 1.1 s) thereafter, and was compared to a single level ATC (sensing level = 50%, tau = 1.4 s). The two ATC circuits were tested with 22 arrhythmia EGMs to determine sensitivity and specificity rates at +/-1-, 2-, 5-, 10-, and 20-mV amplitudes. It was confirmed that a dual level ATC significantly improves the sensitivity rate without degrading the high specificity rate of a standard sensing circuit.


Subject(s)
Arrhythmias, Cardiac/diagnosis , Defibrillators, Implantable , Equipment Design , Humans , Sensitivity and Specificity
9.
Circulation ; 94(7): 1768-73, 1996 Oct 01.
Article in English | MEDLINE | ID: mdl-8840873

ABSTRACT

BACKGROUND: Multiple factors contribute to an improved defibrillation threshold of biphasic shocks. The leading-edge voltage of the second phase may be an important factor in reducing the defibrillation threshold. METHODS AND RESULTS: We tested two experimental biphasic waveforms with large voltage changes at phase reversal. The phase 2 leading-edge voltage was twice the phase 1 trailing-edge voltage. This large voltage change was achieved by switching two capacitors from parallel to series mode at phase reversal. Two capacitors were tested (60/15 microfarads [microF] and 90/22.5 microF) and compared with two control biphasic waveforms for which the phase 1 trailing-edge voltage equaled the phase 2 leading-edge voltage. The control waveforms were incorporated into clinical (135/135 microF) or investigational devices (90/90 microF). Defibrillation threshold parameters were evaluated in eight anesthetized pigs by use of a nonthoracotomy transvenous lead to a can electrode system. The stored energy at the defibrillation threshold (ion joules) was 8.2 +/- 1.5 for 60/15 microF (P < .01 versus 135/135 microF and 90/90 microF), 8.8 +/- 2.4 for 90/22.5 microF (P < .01 versus 135/135 microF and 90/90 microF), 12.5 +/- 3.4 for 135/135 microF, and 12.6 +/- 2.6 for 90/90 microF. CONCLUSIONS: The biphasic waveform with large voltage changes at phase reversal caused by parallel-series mode switching appeared to improve the ventricular defibrillation threshold in a pig model compared with a currently available biphasic waveform. The 60/15-microF capacitor performed as well as the 90/ 22.5-microF capacitor in the experimental waveform. Thus, smaller capacitors may allow reduction in device size without sacrificing defibrillation threshold energy requirements.


Subject(s)
Electric Countershock/methods , Animals , Differential Threshold , Electricity , Equipment Design , Swine , Ventricular Function