Proc Natl Acad Sci U S A ; 104(22): 9187-91, 2007 May 29.
Article in English | MEDLINE | ID: mdl-17488819

ABSTRACT

Silicon micromachined calorimeters ("calorimeter on a chip") are used to measure heat capacities and phase transition enthalpies for thin film, single crystal, and powder samples (5-500 mug). The technology is thus compatible with the small samples produced in multianvil and large diamond anvil cells. Techniques for handling small samples and attaching them to the calorimetric devices have been developed. Initial data illustrate application to CoO and to Fe(2)SiO(4) olivine and spinel, a quenched high pressure phase metastable at ambient conditions. The calorimetric entropy of the olivine-spinel transition in Fe(2)SiO(4) (-16 +/- 5 J/mol.K) is in good agreement with that calculated from phase equilibrium data (-14 +/- 3 J/mol.K). A magnetic transition in iron silicate spinel, detected previously by Mossbauer spectroscopy, is seen in the calorimetric signal.


Subject(s)
Calorimetry/instrumentation , Calorimetry/methods , Cobalt/chemistry , Iron Compounds/chemistry , Minerals/chemistry , Oxides/chemistry , Oxygen/chemistry , Pressure , Silicon/chemistry , Temperature