An existing apparatus for forced flow through of liquid sorbents has been enhanced with an optically accessible system including a transparent crucible, high pressure viewing cell, and camera. With this optical system, the active surface area between gas and liquid can be determined in situ for the first time under industrial process conditions while maintaining the accuracy of a magnetic suspension balance. Additionally, occurring swelling and the resulting buoyancy changes can now be corrected, further improving the quality of the data. Validation measurements focusing on the sorption isotherms, swelling, and bubble geometry of 1-butyl-3-methylimidazolium tetrafluoroborate with nitrogen at 303 K and up to 17 MPa, as well as with carbon dioxide at 303 K, 323 K, and 373 K at up to 3.5 MPa were completed. Absorption of nitrogen resulted in no observable volume change, whereas absorption of carbon dioxide resulted in temperature independent swelling of up to 9.8%. The gas bubble's structure and behavior during its ascend through the liquid was optically tracked in situ. Combining these two data sets with the absorption kinetics forms the basis to determine the measuring system independent mass transfer coefficients, which are applicable in other laboratory scale and industrial processes.