Hip Int ; 22(1): 90-5, 2012.
Article in English | MEDLINE | ID: mdl-22344479

ABSTRACT

The Stryker Revolution(TM) is a new mixing system that employs a high vacuum and a motorised mixing spatula in an effort to reduce cement porosity. We have compared Revolution(TM) with Depuy Cemvac(®), in terms of system reliability and cement porosity. Standardised Simplex P(®) and SmartSet(®) HV cement samples were produced using both mixing systems and analysed using a micro-CT scanner. The overall porosity, number and volume of voids were measured. Void analysis was subdivided into macro-pores (>0.5 mm3) and micro-pores (0.0005-0.5 mm3). Both systems were easy to use and no breakages were encountered. There was no significant difference in overall porosity between Revolution(TM) and Cemvac(®). Revolution(TM) produced over a five-fold decrease in average macro-pore size with medium viscosity cement (p=0.02), but produced a greater number of micro-pores (p<0.01). SmartSet(®) HV specimens had a higher porosity compared to Simplex P(®). This study demonstrated that the Revolution(TM) system was reliable and reduced porosity at least as effectively as the established Cemvac(®) system. The Revolution(TM) produced a greater number of smaller pores and further testing is required to establish if this results in a significant mechanical benefit.


Subject(s)
Bone Cements/chemistry , X-Ray Microtomography/methods , Porosity , Reproducibility of Results , Solutions