1.
Cancer Radiother
; 24(8): 866-869, 2020 Dec.
Article
in English
| MEDLINE
| ID: mdl-33129716
ABSTRACT
Spinal metastasis are a daily challenge in clinical practice. Stereotactic body radiotherapy (SBRT) allows delivery of definitive treatment with excellent long-term control rates. Its implementation needs dedicated devices and day-to-day image-guided radiotherapy (IGRT). The XSight™ spine tracking system, integrates with the CyberKnife® (Accuray™), provides a fiducial-free tracking system for spinal SBRT. We report a rare case of tracking failure during treatment due to the occurrence of a vertebral compression fracture (VCF).