Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
Radiat Prot Dosimetry ; 119(1-4): 368-74, 2006.
Article in English | MEDLINE | ID: mdl-16990348

ABSTRACT

A new system for in vivo dosimetry during radiotherapy has been introduced. Luminescence signals from a small crystal of carbon-doped aluminium oxide (Al2O3:C) are transmitted through an optical fibre cable to an instrument that contains optical filters, a photomultiplier tube and a green (532 nm) laser. The prime output is continuous wave optically stimulated luminescence (CW-OSL) used for the measurement of the integrated dose. We demonstrate a measurement protocol with high reproducibility and improved linearity, which is suitable for clinical dosimetry. A crystal-specific minimum pre-dose is necessary for signal stabilisation. Simple background subtraction only partially removes the residual signal present at long integration times. Instead, the measurement protocol separates the decay curve into three individual components and only the fast and medium components were used.


Subject(s)
Algorithms , Aluminum Oxide/chemistry , Aluminum Oxide/radiation effects , Fiber Optic Technology/instrumentation , Models, Chemical , Thermoluminescent Dosimetry/instrumentation , Thermoluminescent Dosimetry/methods , Computer Simulation , Dose-Response Relationship, Radiation , Equipment Design , Equipment Failure Analysis , Materials Testing , Radiation Dosage , Reproducibility of Results , Sensitivity and Specificity
SELECTION OF CITATIONS
SEARCH DETAIL
...