Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 2 de 2
Filter
Add more filters










Database
Language
Publication year range
1.
J Med Imaging (Bellingham) ; 4(2): 025501, 2017 Apr.
Article in English | MEDLINE | ID: mdl-28491907

ABSTRACT

This study aimed to determine whether a reduction in radiation dose was found for percutaneous coronary interventional (PCI) patients using a cardiac interventional x-ray system with state-of-the-art image enhancement and x-ray optimization, compared to the current generation x-ray system, and to determine the corresponding impact on clinical image quality. Patient procedure dose area product (DAP) and fluoroscopy duration of 131 PCI patient cases from each x-ray system were compared using a Wilcoxon test on median values. Significant reductions in patient dose ([Formula: see text]) were found for the new system with no significant change in fluoroscopy duration ([Formula: see text]); procedure DAP reduced by 64%, fluoroscopy DAP by 51%, and "cine" acquisition DAP by 76%. The image quality of 15 patient angiograms from each x-ray system (30 total) was scored by 75 clinical professionals on a continuous scale for the ability to determine the presence and severity of stenotic lesions; image quality scores were analyzed using a two-sample [Formula: see text]-test. Image quality was reduced by 9% ([Formula: see text]) for the new x-ray system. This demonstrates a substantial reduction in patient dose, from acquisition more than fluoroscopy imaging, with slightly reduced image quality, for the new x-ray system compared to the current generation system.

2.
Br J Radiol ; 89(1067): 20160269, 2016 Nov.
Article in English | MEDLINE | ID: mdl-27610932

ABSTRACT

OBJECTIVES: This study aimed to determine the impact on radiation dose and image quality of a new cardiac interventional X-ray system for trans-catheter aortic valve implantation (TAVI) patients compared to the previously-used cardiac X-ray system. METHODS: Patient dose and image data were retrospectively collected from a Philips AlluraClarity (new) and Siemens Axion Artis (reference) X-ray system. Patient dose area product (DAP) and fluoroscopy duration of 41 patient cases from each X-ray system were compared using a Wilcoxon test. Ten patient aortograms from each X-ray system were scored by 32 observers on a continuous scale to assess the clinical image quality at the given phase of the TAVI procedure. Scores were dichotomised by acceptability and analysed using a Chi-squared test. RESULTS: Significant reductions in patient dose (p << 0.001) were found for the new system with no significant change in fluoroscopy duration (p = 0.052); procedure DAP reduced by 55%, fluoroscopy DAP by 48% and "cine" acquisition DAP by 61%. There was no significant difference between image quality scores of the two X-ray systems (p = 0.06). CONCLUSIONS: The new cardiac X-ray system demonstrated a very significant reduction in patient dose with no loss of clinical image quality. Advances in Knowledge: The huge growth of TAVI may impact on the radiation exposure of cardiac patients and particularly on operators including anaesthetists; cumulative exposure of interventional cardiologists performing high volume TAVI over 30-40 years may be harmful. The Phillips Clarity upgrade including improved image enhancement and optimised X-ray settings significantly reduced radiation without reducing clinically acceptable image quality.


Subject(s)
Radiography, Interventional/instrumentation , Radiography, Thoracic/instrumentation , Transcatheter Aortic Valve Replacement , Fluoroscopy , Humans , Radiation Dosage , Retrospective Studies , Risk Factors , Time Factors
SELECTION OF CITATIONS
SEARCH DETAIL
...