Opt Express ; 18(25): 26163-8, 2010 Dec 06.
Article in English | MEDLINE | ID: mdl-21164965

ABSTRACT

We present time-resolved cyclotron resonance spectra of holes in p-Ge measured during single magnetic field pulses by using a rapid-scanning, fiber-coupled terahertz time-domain spectroscopy system. The key component of the system is a rotating monolithic delay line featuring four helicoid mirror surfaces. It allows measurements of THz spectra at up to 250 Hz repetition rate. Here we show results taken at 150 Hz. In a single 900 ms measurement 135 cyclotron resonance spectra were recorded that fully agree with what is expected from literature.


Subject(s)
Cyclotrons/instrumentation , Terahertz Spectroscopy/instrumentation , Computer-Aided Design , Electromagnetic Fields , Equipment Design , Equipment Failure Analysis