1.
Appl Radiat Isot ; 160: 109113, 2020 Jun.
Article in English | MEDLINE | ID: mdl-32174464

ABSTRACT

A defined-solid-angle proportional counter has been designed for the absolute determination of X-ray emission rates between 2 keV and 10 keV. The main features of the counter are described in this work, and spectra of two radionuclides with low-energy X-ray emissions are presented. The activity concentration of an 55Fe solution was determined by measuring the emission rate of three samples and using tabulated emission probabilities. This activity concentration is compared with the value obtained by liquid scintillation counting, and an excellent agreement has been found.

2.
Appl Radiat Isot ; 155: 108920, 2020 Jan.
Article in English | MEDLINE | ID: mdl-31622844

ABSTRACT

The probabilities of locating peaks with a high relative peak-area uncertainty were determined empirically with nine types of peak-location software used in laboratories engaged in gamma-ray spectrometry measurements. It was found that it is not possible to locate peaks with a probability of 0.95, when they have a relative peak-area uncertainty in excess of 50%. Locating peaks at these relatively high peak-area uncertainties with a probability greater than 0.95 is only possible in the library-driven mode, where the peak positions are supposed a-priori. The deficiencies of the library-driven mode and the possibilities to improve the probabilities of locating peaks are briefly discussed.