J Appl Clin Med Phys ; 4(1): 51-7, 2003.
Article in English | MEDLINE | ID: mdl-12540818

ABSTRACT

This paper describes an innovative method for correctly estimating the effective field size of tangential-breast fields. The method uses an "equivalent triangle" to verify intact breast tangential field monitor-unit settings calculated by a 3D planning system to within 2%. The effects on verification calculations of loss of full scatter due to beam oblique incidence, proximity to field boundaries, and reduced scattering volumes are handled properly. The methodology is validated by comparing calculations performed by the 3D planning system with the respective verification estimates. The accuracy of this technique is established for dose calculations both with and without heterogeneity corrections.


Subject(s)
Breast/anatomy & histology , Imaging, Three-Dimensional/methods , Models, Anatomic , Radiotherapy Planning, Computer-Assisted/methods , Breast/radiation effects , Humans , Radiotherapy, Computer-Assisted/methods , Radiotherapy, Conformal/methods , Scattering, Radiation , Thoracic Wall/anatomy & histology , Thoracic Wall/radiation effects