1.
Comput Methods Biomech Biomed Engin ; 21(5): 399-407, 2018 Apr.
Article in English | MEDLINE | ID: mdl-29996696

ABSTRACT

We aim to maximize the pumping volume of a pulsatile ventricular assist device, where the diaphragm is covered with an endothelial cell layer. These cells are estimated to survive a cyclic strain up to fifteen percent. To increase the pumping volume under this strain constraint we use an approach based on corrugation of the diaphragm in its reference configuration. The paper explains the parametrization scheme for finding corrugation shapes, addresses modeling and evaluation schemes and reports on the results of a parameter study. The results show that corrugated diaphragm shapes are effective for increasing pumping volumes under a strain constraint.


Subject(s)
Diaphragm/anatomy & histology , Heart-Assist Devices , Materials Testing , Pulsatile Flow , Humans , Models, Theoretical , Pressure