Korean J Anesthesiol ; 65(5): 438-41, 2013 Nov.
Article in English | MEDLINE | ID: mdl-24363847

ABSTRACT

BACKGROUND: The aim of the study was to evaluate the leakage of liquid past the cuffs of tracheal tubes in fresh frozen human heads. METHODS: Six truncated fresh frozen heads were used and intubated with 8.0 mm endotracheal tubes. The intracuff pressures tested were 30 and 100 cmH2O. Subsequently, 20 ml of each of two oral antiseptic rinses (0.2% chlorhexidine and octenidine [octenidol®, Schülke & Mayr GmbH, Norderstedt, Germany]) was applied for thirty seconds in the mouth. During the trial, leakage of the cuffs was examined. RESULTS: The sealing between the tracheal cuff and tracheal wall was leakage-proof for all tested intracuff pressures and all tested antiseptic rinses. However, approximately 5.6 ml and 1.8 ml leaked into the esophagus and remained as a cuff-puddle, respectively. CONCLUSIONS: The sealing between an endotracheal tube cuff with an intracuff pressure of 30 cmH2O and the tracheal wall is leakage-proof during oral care with antiseptic rinsing. An increase of intracuff pressure to 100 cmH2O does not appear to be required.