Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
Ann Vasc Surg ; 58: 180-189, 2019 Jul.
Article in English | MEDLINE | ID: mdl-30684621

ABSTRACT

BACKGROUND: To evaluate the early safety and clinical performance of the new low-profile RelayPro Thoracic Stent-Graft System in patients with thoracic aortic disease. METHODS: This was an international, prospective, single-arm study in patients diagnosed with thoracic aorta disease (aneurysm, pseudoaneurysm, dissection, penetrating atherosclerotic ulcer, or intramural hematoma) and treated with a RelayPro stent-graft (in bare stent and/or nonbare stent configurations). The primary endpoints were freedom from aneurysm or dissection-related mortality and stent-graft performance. RESULTS: A total of 31 patients were treated with the RelayPro thoracic stent-graft between 2014 and 2015 at 8 sites in Italy and Spain. Mean age was 72.1 (±10.2) years and 77% were male, 74% with hypertension, and 42% with a history of smoking. Twenty-four (77%) had aneurysms (fusiform in 46%, saccular in 42%, pseudoaneurysm in 12%); 5 (16%) had penetrating atherosclerotic ulcer; and 2 (6%) had chronic Type B dissection. Mean vascular access diameter was 9.1 mm (6-13 mm); 7 patients (23%) had vascular access of 7 mm or less. Technical success was 100% (primary, 90%; assisted primary, 10%). Freedom from aneurysm/dissection-related mortality through 30 days was 100%. Freedom from device-related major adverse events through 30 days was 94%. At 1 year, there was 1 (3%) type Ib and 1 (3%) type II endoleak, 1 (3%) nonaneurysm-related late death, and 1 (3%) secondary intervention (to correct type Ib endoleak). CONCLUSIONS: The RelayPro has a 3-4 French profile reduction to allow endovascular repair of thoracic aortic disease in patients with smaller anatomies. This study shows good initial stent-graft performance and a favorable early safety profile.


Subject(s)
Aorta, Thoracic/surgery , Aortic Diseases/surgery , Blood Vessel Prosthesis Implantation/instrumentation , Blood Vessel Prosthesis , Endovascular Procedures/instrumentation , Stents , Adult , Aged , Aged, 80 and over , Aorta, Thoracic/diagnostic imaging , Aorta, Thoracic/physiopathology , Aortic Diseases/diagnostic imaging , Aortic Diseases/mortality , Aortic Diseases/physiopathology , Aortography/methods , Blood Vessel Prosthesis Implantation/adverse effects , Blood Vessel Prosthesis Implantation/mortality , Computed Tomography Angiography , Endovascular Procedures/adverse effects , Endovascular Procedures/mortality , Female , Humans , Italy , Male , Middle Aged , Postoperative Complications/mortality , Postoperative Complications/surgery , Progression-Free Survival , Prospective Studies , Prosthesis Design , Reoperation , Risk Assessment , Risk Factors , Spain , Time Factors
SELECTION OF CITATIONS
SEARCH DETAIL
...