Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
Article in English | MEDLINE | ID: mdl-18002426

ABSTRACT

The performance assessment of a novel ECG signal processing technology in Fidelity 100 (test) and four modern ECG systems (controls) was conducted. A quantitative evaluation for one control and a test system was done by simultaneous recordings on 54 patients undergoing percutaneous coronary intervention (PCI) and on a biological reference signal from an ECG simulator. A qualitative performance of baseline wander suppression was done on all five systems. The results showed that the Fidelity 100 system provided excellent detection and quantification of transient ischemia and baseline wander suppression.


Subject(s)
Electrocardiography/instrumentation , Electrocardiography/methods , Ischemic Attack, Transient/diagnosis , Myocardial Ischemia , Signal Processing, Computer-Assisted , Algorithms , Artifacts , Data Interpretation, Statistical , Equipment Design , Humans , Ischemic Attack, Transient/pathology , Reproducibility of Results , Software Design , Time Factors
SELECTION OF CITATIONS
SEARCH DETAIL
...