J Clin Microbiol ; 43(5): 2477-80, 2005 May.
Article in English | MEDLINE | ID: mdl-15872288

ABSTRACT

The clinical significance of specimens with low sample-to-cutoff (S/Co) ratios in the Ortho VITROS chemiluminescence assay (CIA) for detection of antibodies to hepatitis C virus (HCV) was evaluated. In one study of 482 CIA-reactive samples, none of the 83 samples with S/Co ratios of < 5 was HCV RNA positive. In a subsequent study, 332 samples with S/Co ratios of between 1 and 20 were tested with the recombinant immunoblot assay (RIBA). None of the 163 samples with S/Co ratios of < 5 was RIBA positive, 83% were RIBA negative, and 28 samples (18%) were RIBA indeterminate. HCV RNA and/or clinical evidence of hepatitis was not found in the 27 indeterminate cases examined. These results show that over 99% of samples with very low S/Co ratios (< or = 5) have no evidence of HCV infection. Therefore, we suggest that the HCV antibody testing algorithm for the VITROS assay might be modified to eliminate supplemental testing of samples with very low S/Co ratios.


Subject(s)
Hepacivirus/immunology , Hepatitis C Antibodies/blood , Hepacivirus/genetics , Hepacivirus/isolation & purification , Hepatitis C/diagnosis , Hepatitis C/immunology , Humans , Luminescent Measurements , RNA, Viral/isolation & purification , Reproducibility of Results