ABSTRACT

This article describes a power-efficient, high dynamic range (DR) incremental ADC (IADC) for wearable biopotential signals recording, where DC and low-frequency disturbances such as electrode offset, 50/60 Hz interference and motion artifact must be tolerated. To achieve a wide DR, the IADC performs a three-step conversion by combining zoom-SAR and extended counting (EC) on top of a second-order incremental delta-sigma modulator (ΔΣM). The hybrid architecture notably reduces the oversampling ratio (OSR) with respect to conventional incremental ΔΣMs, while using the EC further improves the Signal-to-Noise-and-Distortion Ratio (SNDR) by 7.4 to 25.6 dB. Fabricated in a 0.18-µm CMOS technology, the IADC achieves 107.6-dB DR, 104.9-dB peak SNR, and 99.3-dB peak SNDR at 2 kS/s while dissipating 130 µW from 1.8-V (analog) / 1.2-V (digital) supply. This translates to a highly competitive FoMDR of 176.5 dB. The high-DR IADC reduces the gain of the preceding instrumentation amplifier (IA) such that significant DC and low-frequency disturbances can be tolerated. The advantages of high DR have been demonstrated by wearable Electrocardiography (ECG) and Electroencephalography (EEG) recordings under motion artifact.


Subject(s)
Electrocardiography , Signal Processing, Computer-Assisted , Equipment Design , Amplifiers, Electronic , Motion