J Biochem Biophys Methods ; 43(1-3): 87-111, 2000 Jul 05.
Article in English | MEDLINE | ID: mdl-10869669

ABSTRACT

A unique separator was developed which allowed automatic separation and peak collection using semi-preparative supercritical fluid chromatography (SFC). A peak detector switched the effluent between waste and special collection cassettes. Up to 50 mg of various solutes were injected onto a 21-mm I.D. Cyano column. The entire flow path was contained and no aerosols were generated. Collection efficiency was as high as 95%. Peak purity was often greater than 99. 9%. Typical run times were less than 10 min. An analytical SFC was used to screen the performance of a wide range of mobile and stationary phases for the elution of more than 60 miscellaneous small drug compounds. The best 'universal' gradient employed 0.4% isobutyl or isopropylamine dissolved in methanol, then mixed from 5 to 55% into carbon dioxide at 10%/min. Flow rate was 50 ml/min. The analytical SFC was shown to be a good predictor of the semi-prep instrumental performance.


Subject(s)
Chromatography/instrumentation , Chromatography/methods , Peptide Library , 4-Aminobenzoic Acid/chemistry , Antipyrine/isolation & purification , Benzamides/chemistry , Butylamines/chemistry , Carbon Dioxide/chemistry , Combinatorial Chemistry Techniques/methods , Imipramine/isolation & purification , Methanol/chemistry , Procaine/isolation & purification , Propylamines/chemistry , Time Factors , para-Aminobenzoates