Electrophoresis ; 31(16): 2854-7, 2010 Aug.
Article in English | MEDLINE | ID: mdl-20661945

ABSTRACT

Herein, we report a new CE method to measure adenine nucleotides adenosine 5'-triphosphate, adenosine 5'-diphosphate, and adenosine 5'-monophosphate in red blood cells. For this purpose, 20 mmol/L sodium acetate buffer at pH 3.80 was used as running electrolyte, and the separation was performed by the simultaneous application of a CE voltage of 25 kV and an overimposed pressure of 0.2 psi from inlet to outlet. A rapid separation of these analytes in less than 1.5 min was obtained with a good reproducibility for intra- and inter-assay (CV<4 and 8%, respectively) and an excellent analytical recovery (from 98.3 to 99%). The applicability of our method was proved by measuring adenine nucleotides in red blood cells.


Subject(s)
Adenosine Diphosphate/isolation & purification , Adenosine Monophosphate/isolation & purification , Adenosine Triphosphate/isolation & purification , Electrophoresis, Capillary/methods , Adenosine Diphosphate/blood , Adenosine Monophosphate/blood , Adenosine Triphosphate/blood , Electrophoresis, Capillary/instrumentation , Erythrocytes/cytology , Humans , Pressure , Reproducibility of Results , Temperature