Magn Reson Med ; 49(3): 551-7, 2003 Mar.
Article in English | MEDLINE | ID: mdl-12594759

ABSTRACT

A multiple-mouse solenoidal MR coil was developed for in vivo imaging of up to 13 mice simultaneously to screen for tumors on a 1.5 T clinical scanner. For the coil to be effective as a screening tool, it should permit acquisition of MRIs in which orthotopic tumors with diameters >2 mm are detectable in a reasonable period of time (<1 hr magnet time) and their sizes accurately measured. Using a spin echo sequence, we demonstrated that this coil provides sufficient sensitivity for moderately high resolution images (156-176 microm in plane-resolution, 1.5 mm slice thickness). This spatial resolution permitted detection of primary brain tumors in transgenic/knockout mice and orthotopic xenografts. Brain tumor size as measured by MRI was correlated with size measured by histopathology (P < 0.001). Metastatic tumors in the mouse lung were also successfully imaged in a screening setting. The multiple mouse coil is simple in construction and may be implemented without any significant modification to the hardware or software on a clinical scanner.


Subject(s)
Brain Neoplasms/diagnosis , Lung Neoplasms/diagnosis , Magnetic Resonance Imaging/instrumentation , Models, Animal , Animals , Carcinoma, Lewis Lung/diagnosis , Equipment Design , Feasibility Studies , Glioma/diagnosis , Mice , Mice, Knockout , Mice, Transgenic