J Cardiovasc Electrophysiol ; 30(7): 1102-1107, 2019 07.
Article in English | MEDLINE | ID: mdl-30983092

ABSTRACT

Several Boston Scientific pacemaker models have a known issue with intermittent oversensing of the minute ventilation sensor when paired with non-Boston Scientific leads. Several of our patients with these hybrid systems have had transient out of range impedances and oversensing after safety switching which we suspected may be related. A retrospective analysis of 395 patients who had pacemakers implanted between 2015-2017 found that transient out of range impedances with safety switching was present in 9% of Boston Scientific pacemakers paired with Abbott or Medtronic leads compared with 0% in other device-lead combinations (P = 0.0089). We postulate that the root cause of the minute ventilation oversensing and transient high impedance issue is the same, a header-lead interaction from low-level incompatibility. Recognizing this issue is critical to prevent unnecessary lead revisions or extractions as it can be prevented with a simple reprogramming of lead pace/sense configuration.


Subject(s)
Arrhythmias, Cardiac/therapy , Cardiac Pacing, Artificial/adverse effects , Pacemaker, Artificial , Action Potentials , Aged , Arrhythmias, Cardiac/diagnosis , Arrhythmias, Cardiac/physiopathology , Device Removal , Electric Impedance , Equipment Design , Equipment Failure , Female , Heart Rate , Humans , Male , Middle Aged , Retrospective Studies , Risk Factors , Time Factors , Treatment Outcome