Phys Med Biol ; 43(12): 3649-59, 1998 Dec.
Article in English | MEDLINE | ID: mdl-9869039

ABSTRACT

Method validation techniques were developed and experiments were carried out using a beam imaging system (BIS, Wellhöfer Dosimetrie, Schwarzenbruck, Germany) for routine quality assurance of medical accelerators. The routine quality assurance tasks include x-ray beam flatness and symmetry check, light/radiation field congruence test, beam energy constancy for electrons and mechanical checks for couch and collimator rotations. Comparisons were made between the BIS application and conventional quality assurance methods that use radiographic films or detector arrays. In this work, we have demonstrated efficiency and accuracy of the BIS to perform some of the routine quality assurance tasks for medical linear accelerators.


Subject(s)
Particle Accelerators/standards , Radiographic Image Interpretation, Computer-Assisted , Radiotherapy, High-Energy/standards , Electrons , Film Dosimetry , Mathematical Computing , Quality Control , Radiographic Image Interpretation, Computer-Assisted/instrumentation , Radionuclide Imaging , Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted , Scintillation Counting , X-Rays
2.
Med Phys ; 24(8): 1213-20, 1997 Aug.
Article in English | MEDLINE | ID: mdl-9284242

ABSTRACT

A quality assurance procedure was developed for x-ray beam intensity modulated conformal radiotherapy (IMCRT) using dynamic multileaf collimators (MLC). The procedure verifies a prescribed intensity modulated x-ray beam pattern in the beam eye's view (BEV) before the treatment procedure is applied to a patient. It verifies that (a) the leaf sequencing computer files were transferred correctly to the linac control computer; (b) the treatment can be correctly executed without machine faults. A fast beam imaging system (BIS) consisting of a Gd2O2S scintillation screen, a charge-coupled device (CCD) camera, and a portable personal computer (Wellhöfer Dosimetrie, Schwarzenbruck, Germany) was commissioned for this purpose. Measurements for the BIS performance are presented in this work. Reference images were derived from MLC leaf sequencing files that were used to drive a dynamic MLC system (Varian Oncology Systems, Palo Alto, CA). A correlation method was developed to compare the BIS measurements with the calculated reference images. A correlation coefficient calculated using 26 correct intensity modulated fields was shown to be a reliable threshold to identify inaccurate treatment delivery files. The study has demonstrated the feasibility of using the BIS and the correlation method to carry out on-line quality assurance tasks for IMCRT treatment fields in the BEV.


Subject(s)
Radiotherapy, Computer-Assisted/instrumentation , Radiotherapy, Computer-Assisted/standards , Biophysical Phenomena , Biophysics , Evaluation Studies as Topic , Humans , Models, Theoretical , Neoplasms/diagnostic imaging , Neoplasms/radiotherapy , Particle Accelerators , Quality Control , Radiographic Image Enhancement , Radiotherapy Planning, Computer-Assisted/instrumentation , Radiotherapy Planning, Computer-Assisted/standards , Radiotherapy Planning, Computer-Assisted/statistics & numerical data , Radiotherapy, Computer-Assisted/statistics & numerical data , Radiotherapy, High-Energy/instrumentation , Radiotherapy, High-Energy/standards , Radiotherapy, High-Energy/statistics & numerical data