Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 2 de 2
Filter
Add more filters










Database
Language
Publication year range
1.
Rev Sci Instrum ; 89(8): 086103, 2018 Aug.
Article in English | MEDLINE | ID: mdl-30184660

ABSTRACT

The newly designed and constructed electrostatic quadrupole doublet (EQD) at the University of North Texas (UNT) has achieved mass independent focusing of MeV particles to a spot size of 3.3 × 3.5 µm. The EQD is compared to the Louisiana magnetic doublet which is also in use at UNT. Characteristics such as demagnification and sensitivity to aberrations are simulated by the matrix raytracing software, propagation of rays and aberrations by matrices and compared for each focusing system. Particle induced x-ray emission (PIXE) maps of a 2000 mesh Cu grid are compared and show that both doublets produce suitable spot sizes for microprobe analysis. A coarser, 200 mesh grid and incident beams of 2 MeV H+ and O+ show the EQD to be stigmatic given common aperture sizes and lens potentials.

2.
Rev Sci Instrum ; 79(3): 036102, 2008 Mar.
Article in English | MEDLINE | ID: mdl-18377047

ABSTRACT

A high energy focused ion beam microprobe using a doublet arrangement of short magnetic quadrupole lenses was used to focus 1-3 MeV protons to spot sizes of 1x1 microm2 and 1-4.5 MeV carbon and silicon ion beams to spot sizes of 1.5x1.5 microm2. The results presented clearly demonstrate that this simple doublet configuration can provide high energy microbeams for microanalysis and microfabrication applications.