Appl Spectrosc ; 68(1): 69-78, 2014.
Article in English | MEDLINE | ID: mdl-24405956

ABSTRACT

This paper describes the design and setup of a high-throughput Raman system for an array of eight parallel catalytic reactors during reaction conditions. The "operando" methodology combines in situ spectroscopy during catalytic reaction with a simultaneous activity measurement. The high-throughput operando Raman system, multi-operando, is a device that automates this operando methodology for several catalyst samples at the same time, all samples being in the same reaction conditions. We describe how the system is made, how Raman system positions and acquires spectra, and how each reactor outlet gas is selected and analyzed.