Talanta ; 76(1): 107-10, 2008 Jun 30.
Article in English | MEDLINE | ID: mdl-18585249

ABSTRACT

The determination of fluoride ions in water samples is accomplished by using a tubular flow through detector constructed by drilling a channel through a commercially available LaF(3) crystal electrode in such a way that the original contacts of the non-modified unit are maintained. Its performance when incorporated in both FIA and SIA systems was evaluated and the results show that the tubular unit retains the characteristics of the non-modified electrode. In SIA conditions an extended linear range of response and lower detection limit were achieved when compared with the electrode performance in FIA conditions. These aspects together with the additional advantage of low sample and reagent consumptions in SIA when compared to FIA, makes the incorporation of the proposed tubular ISE in a SIA system the preferred approach for on line determination and monitoring of fluoride content in natural water samples.