Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 6 de 6
Filter
Add more filters










Database
Language
Publication year range
1.
IEEE Sens J ; 15(12): 7081-7086, 2015 Dec.
Article in English | MEDLINE | ID: mdl-27087788

ABSTRACT

We present design, implementation, and characterization of a single photon detector based on 32-channel PMT sensor [model H7260-20, Hamamatsu]. The developed high speed electronics enables the photon counting with linear dynamic range (LDR) up to 108count/s per detector's channel. The experimental characterization and Monte-Carlo simulations showed that in the single photon counting mode the LDR of the PMT sensor is limited by (i) "photon" pulse width (current pulse) of 900ps and (ii) substantial decrease of amplitudes of current pulses for count rates exceeding 108 count/s. The multi-channel architecture of the detector and the developed firm/software allow further expansion of the dynamic range of the device by 32-fold by using appropriate beam shaping. The developed single photon counting detector was tested for the detection of fluorescence labeled microbeads in capillary flow.

2.
Biosens Bioelectron ; 39(1): 152-5, 2013 Jan 15.
Article in English | MEDLINE | ID: mdl-22857907

ABSTRACT

Single photon counting is the most sensitive and accurate method for detection of very weak fluorescent signals obtained in many applications such as DNA sequencing, detection of biological reporters on micro-beads, detection of droplets in micro-fluidic systems, etc. In this paper we describe the use of single photon spectrometer for detection and characterization of very weak multicolor fluorescence produced by mixtures of various fluorescent dyes and quantum dots.


Subject(s)
Fluorescent Dyes/analysis , Quantum Dots , DNA/analysis , Photons , Spectrometry, Fluorescence
3.
Biosens Bioelectron ; 21(10): 1924-31, 2006 Apr 15.
Article in English | MEDLINE | ID: mdl-16516459

ABSTRACT

Post-PCR fragment analysis was conducted using our single photon detection-based DNA sequencing instrument in order to substantially enhance the detection of nucleic biomarkers. Telomerase Repeat Amplification Protocol assay was used as a model for real-time PCR-based amplification and detection of DNA. Using TRAPeze XL kit, telomerase-extended DNA fragments were obtained in extracts of serial 10-fold dilutions of telomerase-positive cells, then amplified and detected during 40-cycle real-time PCR. Subsequently, characteristic 6-base DNA ladder patterns were revealed in the post-PCR samples with capillary electrophoresis (CE). In our CE instrument, fluorescently labeled DNA fragments separate in a single-capillary module and are illuminated by a fiberized Ar-ion laser. The laser-induced fluorescence (LIF) is filtered and detected by the fiberized single photon detector (SPD). To assess the sensitivity of our instrument, we performed PCR at fewer cycles (29 and 25), so that the PCR machine could detect amplification only in the most concentrated samples, and then examined samples with CE. Indeed, PCR has detected amplification in samples with minimum 10(4) cells at 29 cycles and over 10(5) cells at 25 cycles. In contrast, the SPD-based CE-LIF has revealed 6-base repeats in samples with as low as 10(2) cells after 29 cycles and 10(3) cells after 25 cycles. Thus, we have demonstrated 100- to 1000-fold increase in the sensitivity of biomarker detection over real-time PCR, making our approach especially suitable for analysis of clinical samples where abundant PCR inhibitors often cause false-negative results.


Subject(s)
Electrophoresis, Capillary , Photons , Polymerase Chain Reaction , Reagent Kits, Diagnostic , Animals , Electrophoresis, Capillary/methods , Humans , Polymerase Chain Reaction/methods
4.
Electrophoresis ; 24(7-8): 1176-83, 2003 Apr.
Article in English | MEDLINE | ID: mdl-12707909

ABSTRACT

We have studied the formation of a resistive region in the capillary during DNA separation. This effect is caused by an unequal change in the mobilities of cations and anions at the interface between the running buffer solution and the capillary. We studied the motion of the resistive region boundary by sequential removal of portions of the affected capillary end. We found that in the process of developing the resistive region the distribution of the electric fields in the capillary changes from uniform to extremely nonuniform, with a very high field (above 1 MV/cm) in the resistive region and a reduced field (80 V/cm) in the rest of the capillary. Though theoretically a resistive region may appear either at the anode (detection) or the cathode (injection) end of the capillary, all previous publications report the formation of the resistive region at the cathode side. In our experiments, however, the anomalous region is formed at the anode. Thus, the separated DNA peaks move towards the slowly progressing resistive region. Our results indicate that the DNA is stopped at the boundary and does not enter the region. When the resistive region is clipped off the peak motion resumes. This suggests that there exists a potential barrier at the resistive layer boundary that prevents the drift of the peaks towards the anode. The formation of the resistive region interferes with a normal separation process causing a gradual decrease of the capillary current and the deceleration and eventual quenching of the peak motion. For the ABI chemistry, we experimented with adding polymers to the electrode buffer to equate the transference numbers for anions and cations, and found the conditions at which this effect is completely eliminated.


Subject(s)
DNA/isolation & purification , Electrophoresis, Capillary/standards , Buffers , Electric Impedance , Electrophoresis, Capillary/methods , Equipment Failure Analysis , Polymers
5.
Electrophoresis ; 24(7-8): 1184-92, 2003 Apr.
Article in English | MEDLINE | ID: mdl-12707910

ABSTRACT

Recently, we developed a family of high-performance automated capillary DNA sequencing instruments based on a single-photon detection of fluorescently labeled DNA fragments. Our machines employ digital and broadband techniques, essential for achieving superior instrument sensitivity and dynamic range. In the present paper, we discuss limitations of the instrument's performance caused by the nonlinearity of single-photon detectors as well as methods for nonlinearity compensation which increase the detection dynamic range and base-calling accuracy.


Subject(s)
Electrophoresis, Capillary/instrumentation , Nucleotides/analysis , Sequence Analysis, DNA/instrumentation , Electronics , Electrophoresis, Capillary/standards , Equipment Design , Fluorescent Dyes/analysis , Photons , Reference Values , Reproducibility of Results , Sequence Analysis, DNA/standards
6.
Electrophoresis ; 23(16): 2804-17, 2002 Aug.
Article in English | MEDLINE | ID: mdl-12210185

ABSTRACT

We have developed a family of high-performance capillary DNA sequencing instruments based on a novel multicolor fluorescent detection technology. This technology is based on two technical innovations: the multilaser excitation of fluorescence of labeled DNA fragments and the "color-blind" single-photon detection of modulated fluorescence. Our machines employ modern digital and broadband techniques that are essential for achieving superior instrument performance. We discuss the design and testing results for several versions of the automated single lane DNA sequencers, as well as our approach to scaling up to multilane instruments.


Subject(s)
Electrophoresis, Capillary/instrumentation , Sequence Analysis, DNA/instrumentation , Color , Equipment Design , Fluorescent Dyes , Lasers , Photons , Sensitivity and Specificity
SELECTION OF CITATIONS