Phys Med Biol ; 63(5): 055001, 2018 02 26.
Article in English | MEDLINE | ID: mdl-29384730

ABSTRACT

This paper describes end-to-end test procedures as the last fundamental step of medical commissioning before starting clinical operation of the MedAustron synchrotron-based pencil beam scanning (PBS) therapy facility with protons. One in-house homogeneous phantom and two anthropomorphic heterogeneous (head and pelvis) phantoms were used for end-to-end tests at MedAustron. The phantoms were equipped with alanine detectors, radiochromic films and ionization chambers. The correction for the 'quenching' effect of alanine pellets was implemented in the Monte Carlo platform of the evaluation version of RayStation TPS. During the end-to-end tests, the phantoms were transferred through the workflow like real patients to simulate the entire clinical workflow: immobilization, imaging, treatment planning and dose delivery. Different clinical scenarios of increasing complexity were simulated: delivery of a single beam, two oblique beams without and with range shifter. In addition to the dose comparison in the plastic phantoms the dose obtained from alanine pellet readings was compared with the dose determined with the Farmer ionization chamber in water. A consistent systematic deviation of about 2% was found between alanine dosimetry and the ionization chamber dosimetry in water and plastic materials. Acceptable agreement of planned and delivered doses was observed together with consistent and reproducible results of the end-to-end testing performed with different dosimetric techniques (alanine detectors, ionization chambers and EBT3 radiochromic films). The results confirmed the adequate implementation and integration of the new PBS technology at MedAustron. This work demonstrates that alanine pellets are suitable detectors for end-to-end tests in proton beam therapy and the developed procedures with customized anthropomorphic phantoms can be used to support implementation of PBS technology in clinical practice.


Subject(s)
Alanine/chemistry , Head/diagnostic imaging , Pelvis/diagnostic imaging , Phantoms, Imaging , Protons , Radiometry/methods , Radiotherapy Planning, Computer-Assisted/methods , Humans , Monte Carlo Method , Radiation Dosage , Synchrotrons
2.
Phys Med Biol ; 62(23): 8832-8849, 2017 Nov 09.
Article in English | MEDLINE | ID: mdl-28984277

ABSTRACT

A UK multicentre audit to evaluate HDR and PDR brachytherapy has been performed using alanine absolute dosimetry. This is the first national UK audit performing an absolute dose measurement at a clinically relevant distance (20 mm) from the source. It was performed in both INTERLACE (a phase III multicentre trial in cervical cancer) and non-INTERLACE brachytherapy centres treating gynaecological tumours. Forty-seven UK centres (including the National Physical Laboratory) were visited. A simulated line source was generated within each centre's treatment planning system and dwell times calculated to deliver 10 Gy at 20 mm from the midpoint of the central dwell (representative of Point A of the Manchester system). The line source was delivered in a water-equivalent plastic phantom (Barts Solid Water) encased in blocks of PMMA (polymethyl methacrylate) and charge measured with an ion chamber at 3 positions (120° apart, 20 mm from the source). Absorbed dose was then measured with alanine at the same positions and averaged to reduce source positional uncertainties. Charge was also measured at 50 mm from the source (representative of Point B of the Manchester system). Source types included 46 HDR and PDR 192Ir sources, (7 Flexisource, 24 mHDR-v2, 12 GammaMed HDR Plus, 2 GammaMed PDR Plus, 1 VS2000) and 1 HDR 60Co source, (Co0.A86). Alanine measurements when compared to the centres' calculated dose showed a mean difference (±SD) of +1.1% (±1.4%) at 20 mm. Differences were also observed between source types and dose calculation algorithm. Ion chamber measurements demonstrated significant discrepancies between the three holes mainly due to positional variation of the source within the catheter (0.4%-4.9% maximum difference between two holes). This comprehensive audit of absolute dose to water from a simulated line source showed all centres could deliver the prescribed dose to within 5% maximum difference between measurement and calculation.


Subject(s)
Brachytherapy , Clinical Audit , Clinical Trials, Phase III as Topic , Radiation Dosage , Algorithms , Catheters , Female , Humans , Iridium Radioisotopes/therapeutic use , Phantoms, Imaging , Radiometry , Radiotherapy Dosage , Uterine Cervical Neoplasms/radiotherapy