A validated approach to determine various methionine cycle metabolites (S-adenosylmethionine, S-adenosylhomocysteine, and methylthioadenosine) in human blood plasma is offered. The approach is based on solid-phase extraction (with grafted phenylboronic acid) and derivatization with chloroacetaldehyde followed by ultra-performance liquid chromatography with fluorescence detection. We used a 100 × 2.1 mm × 1.8 µm C18 column for the selective separation of analytes. Chromatographic separation was achieved with gradient elution of acetonitrile (flow rate 0.2 mL/min) from 2 to 20%. The eluent was initially composed of 10 mM KH2PO4 with 10 mM acetic acid and 25 µM heptafluorobutyric acid. The total analysis time was 11 min. Validation of the method included detection of the limit of detection (2 nM), limit of quantification (5 nM), accuracy (97.2-101%), and intra- and interday precision (2.2-9.0%). Analysis of plasma samples from healthy volunteers revealed that the average levels of S-adenosylmethionine, S-adenosylhomocysteine, and methylthioadenosine were 93.6, 20.9 and 14.8 nM, respectively.


Subject(s)
Chromatography, High Pressure Liquid/methods , S-Adenosylhomocysteine/blood , S-Adenosylmethionine/blood , Chromatography, High Pressure Liquid/instrumentation , Fluorescence , Humans , Plasma/chemistry , S-Adenosylhomocysteine/isolation & purification , S-Adenosylmethionine/isolation & purification , Solid Phase Extraction