1.
J Comput Assist Tomogr ; 18(2): 326-8, 1994.
Article in English | MEDLINE | ID: mdl-8126295

ABSTRACT

OBJECTIVE: We propose a simple, compact, and accurate light guidance system for interventional CT. MATERIALS AND METHODS: The system using intersecting laser beams is mounted on the CT gantry and ensures precise needle guidance even when the gantry is tilted. The device does not require additional software/hardware. RESULTS: Phantom simulations and clinical experiments have demonstrated an accuracy of +/- 1 degrees and +/- 1 mm. CONCLUSION: The laser guidance system for CT makes the intervention safer, faster, and more accurate.


Subject(s)
Lasers , Radiography, Interventional/methods , Tomography, X-Ray Computed/methods , Biopsy, Needle/instrumentation , Biopsy, Needle/methods , Equipment Design , Humans , Models, Structural , Radiography, Interventional/instrumentation , Tomography, X-Ray Computed/instrumentation