ASAIO J ; 44(4): 309-13, 1998.
Article in English | MEDLINE | ID: mdl-9682958

ABSTRACT

Long-term experience with 63 polyurethane, pail handle, coiled tip peritoneal dialysis catheters surgically implanted in 57 consecutive patients with renal failure is presented. One hundred percent follow-up of the study group represented 1,248 patient-months of observation. Cumulative catheter survival rates were 80.8% at 12 months, 62.3% at 24 months, and 48.1% through 51 months. Catheter half-life was 32.6 months. Infection was the most frequent catheter related complication. Incidence rate of peritonitis was 0.73, and exit site/tunnel infection was 0.42 episodes per patient-year. Median time to first episode was 11.7 months for peritonitis, and 26.3 months for exit site/tunnel infection. Infection led to removal of 28.6% of implanted devices, mechanical blockage resulted in 6.4% loss, and pericatheter leak and tubing break each accounted for 1.6% of catheter removals. The polyurethane, pail handle, coiled tip peritoneal catheter was found to be a reliable long-term access device compared with reported performances of other catheter types. An adverse outcome was identified in the current clinical series with a model design using a permanently attached catheter adapter that caused large exit site wounds that were predisposed to infection and catheter loss.


Subject(s)
Biocompatible Materials , Catheters, Indwelling/adverse effects , Peritoneal Dialysis/instrumentation , Peritonitis/etiology , Polyurethanes , Adolescent , Adult , Aged , Female , Follow-Up Studies , Humans , Male , Middle Aged , Peritonitis/microbiology
SELECTION OF CITATIONS
SEARCH DETAIL
...