Adv Space Res ; 30(4): 917-25, 2002.
Article in English | MEDLINE | ID: mdl-12539758

ABSTRACT

The Mobile Radiation Exposure Control System's (Liulin-4 type) main purpose is to monitor simultaneously the doses and fluxes at 4 independent places. It can also be used for personnel dosimetry. The system consists of 4 battery-operated 256-channel dosimeters-spectrometers. We describe results obtained during the calibrations of the spectrometers at the Cyclotron facilities of the University of Louvain, Belgium and of the National Institute of Radiological Sciences-STA, Chiba, Japan with protons of energies up to 70 MeV. The angular sensitivities of the devices are studied and compared with Monte-Carlo predictions. We also present the results obtained at the HIMAC accelerator with 500 MeV/u Fe ions and at the CERN high energy radiation reference fields. Records made during airplane flights are shown and compared with the predictions of the CARI-6 model.


Subject(s)
Aircraft/instrumentation , Heavy Ions , Neutrons , Radiation Monitoring/instrumentation , Aerospace Medicine , Calibration , Cyclotrons , Equipment Design , Models, Theoretical , Monte Carlo Method , Particle Accelerators , Radiation Dosage , Radiometry