J Endourol ; 29(9): 1076-82, 2015 Sep.
Article in English | MEDLINE | ID: mdl-25809547

ABSTRACT

PURPOSE: To evaluate the ignition and burn risk associated with contemporary fiberoptic and distal sensor endoscopic technologies. MATERIALS AND METHODS: We used new and used SCB Xenon 300 light sources to illuminate a 4.8 mm fiberoptic cable, 10 mm laparoscope, 5 mm laparoscope, rigid cystoscope, semirigid ureteroscope, flexible cystoscope, flexible fiberoptic ureteroscope, distal sensor cystoscope, and a distal sensor ureteroscope (Karl Storz, Inc., Tuttlingen, Germany). We measured peak temperatures at the distal end of each device. We then evaluated each device on a flat and folded surgical drape to establish ignition risk. Finally, we evaluated the effects of all devices on human cadaver skin covered by surgical drape. RESULTS: Peak temperatures recorded for each device ranged from 26.9°C (flexible fiberoptic ureteroscope) to 194.5°C (fiberoptic cable). Drape ignition was noted when the fiberoptic cable was placed against a fold of drape. Contact with the fiberoptic cable, 10 mm laparoscope, 5 mm laparoscope, and distal sensor cystoscope resulted in cadaver skin damage. Cadaver skin damage occurred despite little or no visible change to the surgical drape. Rigid and flexible fiberoptic cystoscopes and flexible fiberoptic ureteroscopes had no effect on surgical drapes or cadaver skin. CONCLUSIONS: Fiberoptic light cables and some endoscopic devices have the potential to cause thermal injury and drape ignition. Thermal injury may occur without visible damage to drapes. Surgeons should remain vigilant regarding the risks associated with these devices and take necessary safety precautions to prevent patient injury.


Subject(s)
Burns/etiology , Cystoscopy/methods , Fiber Optic Technology , Laparoscopy/methods , Optical Fibers , Skin/pathology , Ureteroscopes , Burns/prevention & control , Cadaver , Cystoscopes , Hot Temperature , Humans , Laparoscopes , Light , Patient Safety , Temperature
SELECTION OF CITATIONS
SEARCH DETAIL
...