1.
Opt Lett ; 36(6): 981-3, 2011 Mar 15.
Article in English | MEDLINE | ID: mdl-21403749

ABSTRACT

An optical detection setup consisting of a focused laser beam fed into a downstream Fabry-Perot interferometer (FPI) for demodulation of acoustically generated optical phase variations is investigated for its applicability in photoacoustic tomography. The device measures the time derivative of acoustic signals integrated along the beam. Compared to a setup where the detection beam is part of a Mach-Zehnder interferometer, the signal-to-noise ratio of the FPI is lower, but the image quality of the two devices is similar. Using the FPI in a photoacoustic tomograph allows scanning the probe beam around the imaging object without moving the latter.


Subject(s)
Acoustics , Interferometry/methods , Light , Tomography/methods , Optical Phenomena , Pressure , Time Factors